Development of a new primary humidity measurement standard
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F20%3AN0000016" target="_blank" >RIV/00177016:_____/20:N0000016 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00177016:_____/19:N0000015
Výsledek na webu
<a href="https://iopscience.iop.org/article/10.1088/1361-6501/ab4a54" target="_blank" >https://iopscience.iop.org/article/10.1088/1361-6501/ab4a54</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/1361-6501/ab4a54" target="_blank" >10.1088/1361-6501/ab4a54</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Development of a new primary humidity measurement standard
Popis výsledku v původním jazyce
A new primary humidity generator was developed at the National Standards Authority of Ireland (NSAI). The principles of operation and overall design of the generator are discussed. The flow regime within the saturators was analysed using computational fluid dynamics (CFD) to determine if sufficient mixing takes place within the final saturator. This was later investigated by experiment and it was found that sufficient mixing takes place for flow rates up to 3 l min−1 at standard flow conditions of 101 325 Pa and 293.15 K. The uncertainty due to non-ideal saturation at 90 °C dp is discussed. This uncertainty was analysed with respect to the effect that the flow speed and latent heating in the saturators has on the saturation efficiency, separately. The uncertainty associated with saturation efficiency at 90 °C dew point was estimated from the temperature gradients along the saturation path arising due to latent heating effects. The standard uncertainty due to non-ideal saturation was found to be ±3 mK dp.
Název v anglickém jazyce
Development of a new primary humidity measurement standard
Popis výsledku anglicky
A new primary humidity generator was developed at the National Standards Authority of Ireland (NSAI). The principles of operation and overall design of the generator are discussed. The flow regime within the saturators was analysed using computational fluid dynamics (CFD) to determine if sufficient mixing takes place within the final saturator. This was later investigated by experiment and it was found that sufficient mixing takes place for flow rates up to 3 l min−1 at standard flow conditions of 101 325 Pa and 293.15 K. The uncertainty due to non-ideal saturation at 90 °C dp is discussed. This uncertainty was analysed with respect to the effect that the flow speed and latent heating in the saturators has on the saturation efficiency, separately. The uncertainty associated with saturation efficiency at 90 °C dew point was estimated from the temperature gradients along the saturation path arising due to latent heating effects. The standard uncertainty due to non-ideal saturation was found to be ±3 mK dp.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
20303 - Thermodynamics
Návaznosti výsledku
Projekt
<a href="/cs/project/8B16006" target="_blank" >8B16006: Expansion of European research capabilities in humidity measurement</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2020
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Measurement Science and Technology
ISSN
0957-0233
e-ISSN
1361-6501
Svazek periodika
31
Číslo periodika v rámci svazku
2
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
10
Strana od-do
—
Kód UT WoS článku
000499383000001
EID výsledku v databázi Scopus
—