Deliverable D3: Validation of traceability chains for the absorbed dose rate to water
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F23%3AN0000084" target="_blank" >RIV/00177016:_____/23:N0000084 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Deliverable D3: Validation of traceability chains for the absorbed dose rate to water
Popis výsledku v původním jazyce
In this work, a methodology to determine the absorbed dose rate to water distribution from electronic brachytherapy (eBT) units is suggested. It involves measuring directly in a full scatter water tank with small active detectors whose energy response at low X-ray photon energies have been previously characterized. Dose distributions of an INTRABEAM PRS 500 system in two configurations - bare needle and with the spherical applicator of 40 mm diameter - were measured in water using the Exradin A26 ionization chamber and the PTW 60019 microDiamond detector. The continuous energy response of these two detectors in terms of air kerma from 5 keV to 70 keV (eBT energy range) was available from a previous calibration at several X-ray radiation qualities. Here, it is converted to the response in absorbed dose to water using correction factors obtained with Monte Carlo (MC) particle transport simulations. The detectors were modelled using the MCNP code. The air kerma calibration factors were simulated and compared with measurements to refine and validate the models. A series of simulations of the experimental arrangement of the INTRABEAM and the detector were carried out for determining the system specific detector correction factors. These values are distance-dependent due to the steep dose gradient and the remarkable spectral changes around the eBT source in water. The dose rate with distance from the source, the radial and the polar distributions of the INTRABEAM are reported and compared with the manufacturers supplied data. This work is an effort to provide traceability for detectors and measurement procedures for the determination of 3D dose distributions as part of the European joint research project “Primary standards and traceable measurement methods for X-ray emitting electronic brachytherapy devices” (18NRM02 PRISM-eBT).
Název v anglickém jazyce
Deliverable D3: Validation of traceability chains for the absorbed dose rate to water
Popis výsledku anglicky
In this work, a methodology to determine the absorbed dose rate to water distribution from electronic brachytherapy (eBT) units is suggested. It involves measuring directly in a full scatter water tank with small active detectors whose energy response at low X-ray photon energies have been previously characterized. Dose distributions of an INTRABEAM PRS 500 system in two configurations - bare needle and with the spherical applicator of 40 mm diameter - were measured in water using the Exradin A26 ionization chamber and the PTW 60019 microDiamond detector. The continuous energy response of these two detectors in terms of air kerma from 5 keV to 70 keV (eBT energy range) was available from a previous calibration at several X-ray radiation qualities. Here, it is converted to the response in absorbed dose to water using correction factors obtained with Monte Carlo (MC) particle transport simulations. The detectors were modelled using the MCNP code. The air kerma calibration factors were simulated and compared with measurements to refine and validate the models. A series of simulations of the experimental arrangement of the INTRABEAM and the detector were carried out for determining the system specific detector correction factors. These values are distance-dependent due to the steep dose gradient and the remarkable spectral changes around the eBT source in water. The dose rate with distance from the source, the radial and the polar distributions of the INTRABEAM are reported and compared with the manufacturers supplied data. This work is an effort to provide traceability for detectors and measurement procedures for the determination of 3D dose distributions as part of the European joint research project “Primary standards and traceable measurement methods for X-ray emitting electronic brachytherapy devices” (18NRM02 PRISM-eBT).
Klasifikace
Druh
V<sub>souhrn</sub> - Souhrnná výzkumná zpráva
CEP obor
—
OECD FORD obor
10304 - Nuclear physics
Návaznosti výsledku
Projekt
<a href="/cs/project/8B19003" target="_blank" >8B19003: Primary standards and traceable measurement methods for X-ray emitting electronic brachytherapy devices</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2023
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Počet stran výsledku
24
Místo vydání
PTB Branschweig, Německo
Název nakladatele resp. objednatele
EURAMET
Verze
—