Improved calibration method for the extension of the existing magnetic field calibration capability in Helmholtz coils from 3 MHz to 10 MHz or higher
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F24%3AN0000035" target="_blank" >RIV/00177016:_____/24:N0000035 - isvavai.cz</a>
Výsledek na webu
<a href="https://zenodo.org/records/14184933" target="_blank" >https://zenodo.org/records/14184933</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Improved calibration method for the extension of the existing magnetic field calibration capability in Helmholtz coils from 3 MHz to 10 MHz or higher
Popis výsledku v původním jazyce
The research presented in this report includes the design, fabrication and characterisation of the Helmholtz coils setup up to 10 MHz using various methods. For this purpose, a single-turn shielded small loop antenna and commercial magnetic field probes are used and the results are compared. In addition, to use loop antennas as a calibration transfer standard, the magnetic antenna factors obtained by different calibration methods of loop antennas are compared by including the uncertainty budget of each method. As a result of the evaluation, the magnetic antenna factors obtained by the impedance method of small loop antennas were used to calibrate the magnetic field strength indicated in Helmholtz coils.
Název v anglickém jazyce
Improved calibration method for the extension of the existing magnetic field calibration capability in Helmholtz coils from 3 MHz to 10 MHz or higher
Popis výsledku anglicky
The research presented in this report includes the design, fabrication and characterisation of the Helmholtz coils setup up to 10 MHz using various methods. For this purpose, a single-turn shielded small loop antenna and commercial magnetic field probes are used and the results are compared. In addition, to use loop antennas as a calibration transfer standard, the magnetic antenna factors obtained by different calibration methods of loop antennas are compared by including the uncertainty budget of each method. As a result of the evaluation, the magnetic antenna factors obtained by the impedance method of small loop antennas were used to calibrate the magnetic field strength indicated in Helmholtz coils.
Klasifikace
Druh
V<sub>souhrn</sub> - Souhrnná výzkumná zpráva
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
<a href="/cs/project/9B23022" target="_blank" >9B23022: Development of RF and microwave metrology capability II</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2024
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Počet stran výsledku
36
Místo vydání
Gebze, Turecko
Název nakladatele resp. objednatele
TUBITAK UME
Verze
—