Design and development of a portable tuneable radiation source from UV to IR for in situ calibration of radiometers measuring atmospheric aerosol properties
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00177016%3A_____%2F24%3AN0000056" target="_blank" >RIV/00177016:_____/24:N0000056 - isvavai.cz</a>
Výsledek na webu
<a href="https://iopscience.iop.org/article/10.1088/1742-6596/2864/1/012016" target="_blank" >https://iopscience.iop.org/article/10.1088/1742-6596/2864/1/012016</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/1742-6596/2864/1/012016" target="_blank" >10.1088/1742-6596/2864/1/012016</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Design and development of a portable tuneable radiation source from UV to IR for in situ calibration of radiometers measuring atmospheric aerosol properties
Popis výsledku v původním jazyce
The detection of single photons plays an essential role in advancing single-photon science and technologies. Yet, within the visible/near-infrared spectral region, accurate fibre-based optical power measurements at the few-photon level are not yet well-established. In this study, we report on a fibre-based setup, enabling traceable optical power measurements at the few-photon level in this spectral region. The setup was used to calibrate the detection efficiency (DE) of four single-photon avalanche diode (SPAD) detectors. The relative standard uncertainties on the mean DE values obtained from repeat fibre-to-detector couplings ranged from 0.67% to 0.81% (k = 2). However, the relative standard deviation of DE values, which ranged from 1.38% to 3.20% (k = 2), poses a challenge for the metrology of these devices and applications that require high accuracy and repeatability. We investigated the source of these variations by spatially mapping the response of a detector's fibre connector port, using a focused free-space beam, allowing us to estimate the detector's spatial non-uniformity. In addition, we realise a novel calibration approach for fibre-coupled SPADs in a free-space configuration, enabling a direct comparison between the fibre-based setup and the National Physical Laboratory's established free-space facility using a single SPAD. Finally, we investigated alternative coupling methods, testing the repeatability of different fibre-to-fibre connectors in addition to direct fibre-to-detector couplings: SPADs from three manufacturers were tested, with both single-mode and multi-mode fibre.
Název v anglickém jazyce
Design and development of a portable tuneable radiation source from UV to IR for in situ calibration of radiometers measuring atmospheric aerosol properties
Popis výsledku anglicky
The detection of single photons plays an essential role in advancing single-photon science and technologies. Yet, within the visible/near-infrared spectral region, accurate fibre-based optical power measurements at the few-photon level are not yet well-established. In this study, we report on a fibre-based setup, enabling traceable optical power measurements at the few-photon level in this spectral region. The setup was used to calibrate the detection efficiency (DE) of four single-photon avalanche diode (SPAD) detectors. The relative standard uncertainties on the mean DE values obtained from repeat fibre-to-detector couplings ranged from 0.67% to 0.81% (k = 2). However, the relative standard deviation of DE values, which ranged from 1.38% to 3.20% (k = 2), poses a challenge for the metrology of these devices and applications that require high accuracy and repeatability. We investigated the source of these variations by spatially mapping the response of a detector's fibre connector port, using a focused free-space beam, allowing us to estimate the detector's spatial non-uniformity. In addition, we realise a novel calibration approach for fibre-coupled SPADs in a free-space configuration, enabling a direct comparison between the fibre-based setup and the National Physical Laboratory's established free-space facility using a single SPAD. Finally, we investigated alternative coupling methods, testing the repeatability of different fibre-to-fibre connectors in addition to direct fibre-to-detector couplings: SPADs from three manufacturers were tested, with both single-mode and multi-mode fibre.
Klasifikace
Druh
J<sub>SC</sub> - Článek v periodiku v databázi SCOPUS
CEP obor
—
OECD FORD obor
10509 - Meteorology and atmospheric sciences
Návaznosti výsledku
Projekt
<a href="/cs/project/8B20008" target="_blank" >8B20008: Metrology for aerosol optical properties</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2024
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Physics: Conference Series
ISSN
17426588
e-ISSN
—
Svazek periodika
2864
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
6
Strana od-do
—
Kód UT WoS článku
—
EID výsledku v databázi Scopus
2-s2.0-85208610027