Popis výsledku v původním jazyce
In this study, skin dosimetry of patients undergoing interventional cardiology procedures is presented. Three hospitals were included. Two methods were used for skin dosimetry-radiochromic dosimetry films and reconstruction of skin dose distribution based on examination protocol. Maximum skin doses (MSD) obtained from both methods were compared for 175 patients. For patients for whom the film MSD was } 1 Gy, the reconstruction MSD differed from the film MSD in the range of +/- 50 % for 83 % of patients.For remaining patients, the difference was higher and it was caused by longer fluoroscopy time. For 59 patients for whom the cumulative dose was known, the cumulative dose was compared with the film MSD. Skin dosimetry with radiochromic films is more accurate than the reconstruction method, but films do not include X-ray fields from lateral projections whilest reconstructions do.
Název v anglickém jazyce
Patient skin dosimetry in interventional cardiology in the Czech Republic
Popis výsledku anglicky
In this study, skin dosimetry of patients undergoing interventional cardiology procedures is presented. Three hospitals were included. Two methods were used for skin dosimetry-radiochromic dosimetry films and reconstruction of skin dose distribution based on examination protocol. Maximum skin doses (MSD) obtained from both methods were compared for 175 patients. For patients for whom the film MSD was } 1 Gy, the reconstruction MSD differed from the film MSD in the range of +/- 50 % for 83 % of patients.For remaining patients, the difference was higher and it was caused by longer fluoroscopy time. For 59 patients for whom the cumulative dose was known, the cumulative dose was compared with the film MSD. Skin dosimetry with radiochromic films is more accurate than the reconstruction method, but films do not include X-ray fields from lateral projections whilest reconstructions do.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FA - Kardiovaskulární nemoci včetně kardiochirurgie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/JC_4%2F2008" target="_blank" >JC_4/2008: Radiační zátěž pacientů v intervenční radiologii.</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Protection Dosimetry
ISSN
0144-8420
e-ISSN
—
Svazek periodika
147
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
5
Strana od-do
106-110
Kód UT WoS článku
000296355500025
EID výsledku v databázi Scopus
—