Fixation of the Ajust minisling based upon cadaveric study
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11110%2F13%3A10191755" target="_blank" >RIV/00216208:11110/13:10191755 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00064165:_____/13:10191755
Výsledek na webu
<a href="http://dx.doi.org/10.1007/s00192-013-2139-2" target="_blank" >http://dx.doi.org/10.1007/s00192-013-2139-2</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/s00192-013-2139-2" target="_blank" >10.1007/s00192-013-2139-2</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Fixation of the Ajust minisling based upon cadaveric study
Popis výsledku v původním jazyce
Introduction and hypothesis The objective was to describe the fixation site of the anchor of the Ajust mid-urethral minisling. Methods This cadaveric study was based on a group of 11 formalin-embalmed bodies with legs positioned in 30 degrees flexion and30 degrees abduction, and a group of five fresh-frozen bodies with legs positioned as normal during the procedure. The groups were later compared. The fixation site was dissected and described. The distance to the obturator bundle was considered as theprimary safety parameter. To compare the groups of fresh-frozen bodies and formalin-embalmed bodies, the Student's t test and Mann-Whitney test were used. Results In the group of formalin-embalmed bodies the mean distance from the anchoring device to theobturator nerve was 4.23 cm. In 19 cases out of 22 the anchor was within the complex of the obturator membrane and obturator muscles. In the group of fresh frozen bodies the mean distance to the obturator nerve was 3.15 cm. In 9 cases ou
Název v anglickém jazyce
Fixation of the Ajust minisling based upon cadaveric study
Popis výsledku anglicky
Introduction and hypothesis The objective was to describe the fixation site of the anchor of the Ajust mid-urethral minisling. Methods This cadaveric study was based on a group of 11 formalin-embalmed bodies with legs positioned in 30 degrees flexion and30 degrees abduction, and a group of five fresh-frozen bodies with legs positioned as normal during the procedure. The groups were later compared. The fixation site was dissected and described. The distance to the obturator bundle was considered as theprimary safety parameter. To compare the groups of fresh-frozen bodies and formalin-embalmed bodies, the Student's t test and Mann-Whitney test were used. Results In the group of formalin-embalmed bodies the mean distance from the anchoring device to theobturator nerve was 4.23 cm. In 19 cases out of 22 the anchor was within the complex of the obturator membrane and obturator muscles. In the group of fresh frozen bodies the mean distance to the obturator nerve was 3.15 cm. In 9 cases ou
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FK - Gynekologie a porodnictví
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/NT14162" target="_blank" >NT14162: Porovnání bezpečnosti a perio-operačních komplikací transobturatorně zavedené pásky TVT-O s kotvenou minipáskou s nastavitelnou délkou (Ajust) v randomizované studii</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
International Urogynecology Journal
ISSN
0937-3462
e-ISSN
—
Svazek periodika
24
Číslo periodika v rámci svazku
12
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
5
Strana od-do
2119-2123
Kód UT WoS článku
000328132300020
EID výsledku v databázi Scopus
—