Long-term stability of the Leksell Gamma Knife((R)) Perfexion (TM) patient positioning system (PPS)
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11110%2F14%3A10290740" target="_blank" >RIV/00216208:11110/14:10290740 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68407700:21340/14:00228872 RIV/00064203:_____/14:10294382
Výsledek na webu
<a href="http://dx.doi.org/10.1118/1.4866225" target="_blank" >http://dx.doi.org/10.1118/1.4866225</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1118/1.4866225" target="_blank" >10.1118/1.4866225</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Long-term stability of the Leksell Gamma Knife((R)) Perfexion (TM) patient positioning system (PPS)
Popis výsledku v původním jazyce
Purpose: To assess the long-term mechanical stability and accuracy of the patient positioning system (PPS) of the Leksell Gamma Knife (R) Perfexion (TM) (LGK PFX). Methods: The mechanical stability of the PPS of the LGK PFX was evaluated using measurements obtained between September 2007 and June 2011. Three methods were employed to measure the deviation of the coincidence of the radiological focus point (RFP) and the PPS calibration center point (CCP). In the first method, the onsite diode test tool with single diode detector was used together with the 4 mm collimator on a daily basis. In the second method, a service diode test tool with three diode detectors was used biannually at the time of the routine preventive maintenance. The test performed with the service diode test tool measured the deviations for all three collimators 4, 8, and 16 mm and also for three different positions of the PPS. The third method employed the conventional film pin-prick method. This test was performed a
Název v anglickém jazyce
Long-term stability of the Leksell Gamma Knife((R)) Perfexion (TM) patient positioning system (PPS)
Popis výsledku anglicky
Purpose: To assess the long-term mechanical stability and accuracy of the patient positioning system (PPS) of the Leksell Gamma Knife (R) Perfexion (TM) (LGK PFX). Methods: The mechanical stability of the PPS of the LGK PFX was evaluated using measurements obtained between September 2007 and June 2011. Three methods were employed to measure the deviation of the coincidence of the radiological focus point (RFP) and the PPS calibration center point (CCP). In the first method, the onsite diode test tool with single diode detector was used together with the 4 mm collimator on a daily basis. In the second method, a service diode test tool with three diode detectors was used biannually at the time of the routine preventive maintenance. The test performed with the service diode test tool measured the deviations for all three collimators 4, 8, and 16 mm and also for three different positions of the PPS. The third method employed the conventional film pin-prick method. This test was performed a
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BO - Biofyzika
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Ostatní
Rok uplatnění
2014
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Medical Physics
ISSN
0094-2405
e-ISSN
—
Svazek periodika
41
Číslo periodika v rámci svazku
3
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
9
Strana od-do
—
Kód UT WoS článku
000332485600013
EID výsledku v databázi Scopus
—