USB 3.0 readout and time-walk correction method for Timepix3 detector
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11110%2F16%3A10337874" target="_blank" >RIV/00216208:11110/16:10337874 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1088/1748-0221/11/12/C12065" target="_blank" >http://dx.doi.org/10.1088/1748-0221/11/12/C12065</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1088/1748-0221/11/12/C12065" target="_blank" >10.1088/1748-0221/11/12/C12065</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
USB 3.0 readout and time-walk correction method for Timepix3 detector
Popis výsledku v původním jazyce
The hybrid particle counting pixel detectors of Medipix family are well known. In this contribution we present new USB 3.0 based interface AdvaDAQ for Timepix3 detector. The AdvaDAQ interface is designed with a maximal emphasis to the flexibility. It is successor of FitPIX interface developed in IEAP CTU in Prague. Its modular architecture supports all Medipix/Timepix chips and all their different readout modes: Medipix2, Timepix (serial and parallel), Medipix3 and Timepix3. The high bandwidth of USB 3.0 permits readout of 1700 full frames per second with Timepix or 8 channel data acquisition from Timepix3 at frequency of 320 MHz. The control and data acquisition is integrated in a multiplatform PiXet software (MS Windows, Mac OS, Linux). In the second part of the publication a new method for correction of the time-walk effect in Timepix3 is described. Moreover, a fully spectroscopic X-ray imaging with Timepix3 detector operated in the ToT mode ( Time-over-Threshold) is presented. It is shown that the AdvaDAQ's readout speed is sufficient to perform spectroscopic measurement at full intensity of radiographic setups equipped with nano- or micro-focus X-ray tubes.
Název v anglickém jazyce
USB 3.0 readout and time-walk correction method for Timepix3 detector
Popis výsledku anglicky
The hybrid particle counting pixel detectors of Medipix family are well known. In this contribution we present new USB 3.0 based interface AdvaDAQ for Timepix3 detector. The AdvaDAQ interface is designed with a maximal emphasis to the flexibility. It is successor of FitPIX interface developed in IEAP CTU in Prague. Its modular architecture supports all Medipix/Timepix chips and all their different readout modes: Medipix2, Timepix (serial and parallel), Medipix3 and Timepix3. The high bandwidth of USB 3.0 permits readout of 1700 full frames per second with Timepix or 8 channel data acquisition from Timepix3 at frequency of 320 MHz. The control and data acquisition is integrated in a multiplatform PiXet software (MS Windows, Mac OS, Linux). In the second part of the publication a new method for correction of the time-walk effect in Timepix3 is described. Moreover, a fully spectroscopic X-ray imaging with Timepix3 detector operated in the ToT mode ( Time-over-Threshold) is presented. It is shown that the AdvaDAQ's readout speed is sufficient to perform spectroscopic measurement at full intensity of radiographic setups equipped with nano- or micro-focus X-ray tubes.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Instrumentation
ISSN
1748-0221
e-ISSN
—
Svazek periodika
11
Číslo periodika v rámci svazku
December
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
10
Strana od-do
—
Kód UT WoS článku
000395731900065
EID výsledku v databázi Scopus
2-s2.0-85008155536