Fallacies and Pitfalls of Dialysis Sodium Prescription and Control
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11110%2F18%3A10376733" target="_blank" >RIV/00216208:11110/18:10376733 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00064165:_____/18:10376733
Výsledek na webu
<a href="https://doi.org/10.1159/000484921" target="_blank" >https://doi.org/10.1159/000484921</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1159/000484921" target="_blank" >10.1159/000484921</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Fallacies and Pitfalls of Dialysis Sodium Prescription and Control
Popis výsledku v původním jazyce
Dialysate conductivity is routinely used as a surrogate for dialysate sodium concentration. However, dialysis machine manufacturers apply different conductivity temperature correction coefficients. With the same conductivity in dialysis machines manufactured by different manufacturers, dialysate sodium may significantly differ. Also, electrolyte prescriptions are individualized (K, Ca, HCO3) and this is associated with another variation in dialysate sodium in the order of 1-5 mmol/L and both deviations are cumulative and chronic for each patient. Equivalence of the prescribed dialysate sodium and the concentration measured in it is not granted. Both variables differ and it is machine dependent! This paper analyses those variations from a technical point of view and suggests how to detect them and how to deal with or avoid them in clinical practice.
Název v anglickém jazyce
Fallacies and Pitfalls of Dialysis Sodium Prescription and Control
Popis výsledku anglicky
Dialysate conductivity is routinely used as a surrogate for dialysate sodium concentration. However, dialysis machine manufacturers apply different conductivity temperature correction coefficients. With the same conductivity in dialysis machines manufactured by different manufacturers, dialysate sodium may significantly differ. Also, electrolyte prescriptions are individualized (K, Ca, HCO3) and this is associated with another variation in dialysate sodium in the order of 1-5 mmol/L and both deviations are cumulative and chronic for each patient. Equivalence of the prescribed dialysate sodium and the concentration measured in it is not granted. Both variables differ and it is machine dependent! This paper analyses those variations from a technical point of view and suggests how to detect them and how to deal with or avoid them in clinical practice.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
30217 - Urology and nephrology
Návaznosti výsledku
Projekt
—
Návaznosti
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Ostatní
Rok uplatnění
2018
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Blood Purification
ISSN
0253-5068
e-ISSN
—
Svazek periodika
46
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
CH - Švýcarská konfederace
Počet stran výsledku
7
Strana od-do
27-33
Kód UT WoS článku
000431533100006
EID výsledku v databázi Scopus
2-s2.0-85045312965