Is the fixation of single incision TVT-S tape as good as that of transobturator tape? An ultrasound study of randomized trial results
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11110%2F21%3A10419892" target="_blank" >RIV/00216208:11110/21:10419892 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00064165:_____/21:10419892
Výsledek na webu
<a href="https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=KyAhYpYg9H" target="_blank" >https://verso.is.cuni.cz/pub/verso.fpl?fname=obd_publikace_handle&handle=KyAhYpYg9H</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/s00192-020-04390-6" target="_blank" >10.1007/s00192-020-04390-6</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Is the fixation of single incision TVT-S tape as good as that of transobturator tape? An ultrasound study of randomized trial results
Popis výsledku v původním jazyce
Introduction: TVT-Secur (TVT-S) was the first single-incision sling available on the market and was soon found to have less efficacy than mid-urethral slings. Our aim was to assess the position and tape descent following TVT-O and TVT-S H in a U procedure and, based on this evaluation, to find the possible reason for lower TVT-S surgery efficacy. Methods: We conducted an ultrasound study of a randomised trial with a 3-year follow-up that took place between 2007 and 2009 and included 197 women with urodynamic stress urinary incontinence. Of these, 67 were allocated to receive the TVT-O procedure, 64 to TVT-S in the H position, and 65 in the U position. Patients underwent a complete urogynaecological and ultrasound examination. The positions of the bladder neck and the tape after surgery were assessed and the data obtained compared between groups. Results: Our primary study showed a significantly higher rate of positive stress tests in the TVT-S groups compared to the TVT-O group. After surgery, there was no difference between the position of the tape at rest and at maximal Valsalva between the groups. In TVT-O patients, the mean length of the upper tape margin descent increased from 6 mm on the 1st day after surgery to 9 mm 3 months after surgery and remained stable afterwards. For TVT-S patients, there was a further increase of up to 15 mm in tape descent after 3 months. Conclusions: Our results show that the lower efficacy of TVT-S might be due to inadequate fixation and increasing tape descent.
Název v anglickém jazyce
Is the fixation of single incision TVT-S tape as good as that of transobturator tape? An ultrasound study of randomized trial results
Popis výsledku anglicky
Introduction: TVT-Secur (TVT-S) was the first single-incision sling available on the market and was soon found to have less efficacy than mid-urethral slings. Our aim was to assess the position and tape descent following TVT-O and TVT-S H in a U procedure and, based on this evaluation, to find the possible reason for lower TVT-S surgery efficacy. Methods: We conducted an ultrasound study of a randomised trial with a 3-year follow-up that took place between 2007 and 2009 and included 197 women with urodynamic stress urinary incontinence. Of these, 67 were allocated to receive the TVT-O procedure, 64 to TVT-S in the H position, and 65 in the U position. Patients underwent a complete urogynaecological and ultrasound examination. The positions of the bladder neck and the tape after surgery were assessed and the data obtained compared between groups. Results: Our primary study showed a significantly higher rate of positive stress tests in the TVT-S groups compared to the TVT-O group. After surgery, there was no difference between the position of the tape at rest and at maximal Valsalva between the groups. In TVT-O patients, the mean length of the upper tape margin descent increased from 6 mm on the 1st day after surgery to 9 mm 3 months after surgery and remained stable afterwards. For TVT-S patients, there was a further increase of up to 15 mm in tape descent after 3 months. Conclusions: Our results show that the lower efficacy of TVT-S might be due to inadequate fixation and increasing tape descent.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
30214 - Obstetrics and gynaecology
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2021
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
International Urogynecology Journal
ISSN
0937-3462
e-ISSN
—
Svazek periodika
32
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
8
Strana od-do
141-148
Kód UT WoS článku
000543023100001
EID výsledku v databázi Scopus
2-s2.0-85087073557