AlphaCor artificial cornea: clinical outcome
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11150%2F11%3A10104190" target="_blank" >RIV/00216208:11150/11:10104190 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00179906:_____/11:10104190
Výsledek na webu
<a href="http://www.nature.com/eye/journal/v25/n9/full/eye2011122a.html" target="_blank" >http://www.nature.com/eye/journal/v25/n9/full/eye2011122a.html</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1038/eye.2011.122" target="_blank" >10.1038/eye.2011.122</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
AlphaCor artificial cornea: clinical outcome
Popis výsledku v původním jazyce
The long-term results of AlphaCor implantations, and evaluation of the main complications and risk factors were descrited. The survival rate of the device at 1, 2, and 3 years was 87%, 58%, and 42%, respectively. The most significant complications were stromal melt (nine cases), optic deposition (three eyes), and retroprosthetic membrane formation (three eyes). The most common device-unrelated complication was trauma (three patients). All complications were managed without loss of the eye. AlphaCor provides a treatment option for patients with corneal blindness in which a donor tissue graft would not succeed.
Název v anglickém jazyce
AlphaCor artificial cornea: clinical outcome
Popis výsledku anglicky
The long-term results of AlphaCor implantations, and evaluation of the main complications and risk factors were descrited. The survival rate of the device at 1, 2, and 3 years was 87%, 58%, and 42%, respectively. The most significant complications were stromal melt (nine cases), optic deposition (three eyes), and retroprosthetic membrane formation (three eyes). The most common device-unrelated complication was trauma (three patients). All complications were managed without loss of the eye. AlphaCor provides a treatment option for patients with corneal blindness in which a donor tissue graft would not succeed.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
FF - ORL, oftalmologie, stomatologie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Eye
ISSN
0950-222X
e-ISSN
—
Svazek periodika
25
Číslo periodika v rámci svazku
9
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
9
Strana od-do
1138-1146
Kód UT WoS článku
000294760300006
EID výsledku v databázi Scopus
—