Determination of relative enantiomer migration order using a racemic sample
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11310%2F15%3A10315456" target="_blank" >RIV/00216208:11310/15:10315456 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.chroma.2015.10.058" target="_blank" >http://dx.doi.org/10.1016/j.chroma.2015.10.058</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.chroma.2015.10.058" target="_blank" >10.1016/j.chroma.2015.10.058</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Determination of relative enantiomer migration order using a racemic sample
Popis výsledku v původním jazyce
We developed a method that enables us to distinguish between the same or the opposite enantiomer migration order (EMO) of two enantiomers of a chiral compound with two different selectors. The method is applicable to racemic samples and thus a standard of the pure enantiomeric form(s) is not required. First, complexation constants and mobilities of complexes of the two enantiomers with the first and second selector are determined. However, for a racemic sample it is not possible to deduce whether the first migrating enantiomer with one selector is the same one as the first migrating enantiomer with the second selector. A specific mixture of the two selectors is designed to resolve this. In case the two enantiomers exhibit the same, respectively the opposite EMO in the two selectors, the mixture does, respectively does not separate the racemic sample. Thus two peaks are detected in the first case, while a single coalescent peak is recorded in the opposite case. We demonstrate the method
Název v anglickém jazyce
Determination of relative enantiomer migration order using a racemic sample
Popis výsledku anglicky
We developed a method that enables us to distinguish between the same or the opposite enantiomer migration order (EMO) of two enantiomers of a chiral compound with two different selectors. The method is applicable to racemic samples and thus a standard of the pure enantiomeric form(s) is not required. First, complexation constants and mobilities of complexes of the two enantiomers with the first and second selector are determined. However, for a racemic sample it is not possible to deduce whether the first migrating enantiomer with one selector is the same one as the first migrating enantiomer with the second selector. A specific mixture of the two selectors is designed to resolve this. In case the two enantiomers exhibit the same, respectively the opposite EMO in the two selectors, the mixture does, respectively does not separate the racemic sample. Thus two peaks are detected in the first case, while a single coalescent peak is recorded in the opposite case. We demonstrate the method
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GJ15-18424Y" target="_blank" >GJ15-18424Y: Vývoj nové generace software a pokročilé teorie elektromigrace v metodách elektrokinetické chromatografie</a><br>
Návaznosti
S - Specificky vyzkum na vysokych skolach<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Chromatography A
ISSN
0021-9673
e-ISSN
—
Svazek periodika
1424
Číslo periodika v rámci svazku
neuveden
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
5
Strana od-do
139-143
Kód UT WoS článku
000366772800017
EID výsledku v databázi Scopus
2-s2.0-84948753058