UV-photochemical vapor generation with quartz furnace atomic absorption spectrometry for simple and sensitive determination of selenium in dietary supplements
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216208%3A11310%2F16%3A10318273" target="_blank" >RIV/00216208:11310/16:10318273 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.microc.2015.10.004" target="_blank" >http://dx.doi.org/10.1016/j.microc.2015.10.004</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.microc.2015.10.004" target="_blank" >10.1016/j.microc.2015.10.004</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
UV-photochemical vapor generation with quartz furnace atomic absorption spectrometry for simple and sensitive determination of selenium in dietary supplements
Popis výsledku v původním jazyce
This study describes the optimization and quantification of UV-photochemical vapor generation using the Se(IV) standard, together with application of the proposed technique for practical purposes. The generation process was performed in continuous mode and was combined with atomic absorption spectrometry utilizing an externally heated quartz furnace as an atomizer. Formic acid with a concentration of 0.5 mol L-1 acted as the photochemical agent during the UV-assisted conversion of Se(IV) from the condensed to the gaseous phase; no further chemicals were needed. A very low limit of detection (LOD) 40 ng L-1 was achieved under the optimum experimental conditions and the repeatability expressed as % RSD was 1.8% at 5 mu g L-1 Se(IV). Analysis of the certified reference NIST SRM 1643e was performed to demonstrate the accuracy of the method with good correlation between the certified and experimentally achieved value. The effect of various coexisting ions on the selenium signal was also inv
Název v anglickém jazyce
UV-photochemical vapor generation with quartz furnace atomic absorption spectrometry for simple and sensitive determination of selenium in dietary supplements
Popis výsledku anglicky
This study describes the optimization and quantification of UV-photochemical vapor generation using the Se(IV) standard, together with application of the proposed technique for practical purposes. The generation process was performed in continuous mode and was combined with atomic absorption spectrometry utilizing an externally heated quartz furnace as an atomizer. Formic acid with a concentration of 0.5 mol L-1 acted as the photochemical agent during the UV-assisted conversion of Se(IV) from the condensed to the gaseous phase; no further chemicals were needed. A very low limit of detection (LOD) 40 ng L-1 was achieved under the optimum experimental conditions and the repeatability expressed as % RSD was 1.8% at 5 mu g L-1 Se(IV). Analysis of the certified reference NIST SRM 1643e was performed to demonstrate the accuracy of the method with good correlation between the certified and experimentally achieved value. The effect of various coexisting ions on the selenium signal was also inv
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach<br>I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Microchemical Journal
ISSN
0026-265X
e-ISSN
—
Svazek periodika
124
Číslo periodika v rámci svazku
January
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
10
Strana od-do
584-593
Kód UT WoS článku
000367755600070
EID výsledku v databázi Scopus
—