Popis výsledku v původním jazyce
This publication describes the combination of flow-through electrochemical generation (EcVG) of cadmium volatile form with its in situ trapping in a graphite furnace atomizer. Three cathode materials (Pt, Pb, and Ti) and four potentially suitable electrolytes (HCl, H2SO4, HCOOH and NaCl) were tested. Automated sampling equipment for the graphite furnace atomizer with an untreated fused silica capillary was used for the introduction of the cadmium volatile form into the iridium-treated graphite furnace. The limit of detection (LOD) of the electrochemical Cd volatile form generation with in situ collection was 1.0 ng ml(-1) (concentration LOD) or 1.5 ng (absolute LOD). The efficiency of the method was estimated and discussed. The effect of selected concomitant ions was evaluated and the accuracy of the proposed method was established by determination of the Cd content in the NIST SRM 1643e certified reference material.
Název v anglickém jazyce
Electrochemical generation of volatile form of cadmium and its in situ trapping in a graphite furnace
Popis výsledku anglicky
This publication describes the combination of flow-through electrochemical generation (EcVG) of cadmium volatile form with its in situ trapping in a graphite furnace atomizer. Three cathode materials (Pt, Pb, and Ti) and four potentially suitable electrolytes (HCl, H2SO4, HCOOH and NaCl) were tested. Automated sampling equipment for the graphite furnace atomizer with an untreated fused silica capillary was used for the introduction of the cadmium volatile form into the iridium-treated graphite furnace. The limit of detection (LOD) of the electrochemical Cd volatile form generation with in situ collection was 1.0 ng ml(-1) (concentration LOD) or 1.5 ng (absolute LOD). The efficiency of the method was estimated and discussed. The effect of selected concomitant ions was evaluated and the accuracy of the proposed method was established by determination of the Cd content in the NIST SRM 1643e certified reference material.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA14-23532S" target="_blank" >GA14-23532S: Generování a prekoncentrace těkavých sloučenin pro atomovou absorpční a atomovou fluorescenční spektrometrii</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2016
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Spectrochimica Acta - Part B: Atomic Spectroscopy
ISSN
0584-8547
e-ISSN
—
Svazek periodika
117
Číslo periodika v rámci svazku
March
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
7
Strana od-do
42-48
Kód UT WoS článku
000372385800006
EID výsledku v databázi Scopus
2-s2.0-84957068710