Partially covered, self-expandable metal stents with anti-reflux valves for malignancies comprising hilar region
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14110%2F15%3A00084697" target="_blank" >RIV/00216224:14110/15:00084697 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Partially covered, self-expandable metal stents with anti-reflux valves for malignancies comprising hilar region
Popis výsledku v původním jazyce
Objective: To prove feasibility of percutaneous parallel implantation of partially covered, self-expandable metal stents (PC-SEMS) with antireflux valves (AR valves) in in-vivo model. Method: The novel design of PC-SEMS suitable for patients with hilar malignancies was developed. For percutaneous implantation of stent with AR valve, specific type of introducer is needed to prevent closing and crumpling of valve during implantation procedure. Stent is half covered with e-PTFE membrane, also forming AR valve. Proximal half of the stent is bare to allow flow of bile from segmental branches of biliary tree arising in hilar region. In-vivo experiment was chosen to test safe deployment of two PC-SEMS with AR valve. Parallel and synchronous implantation of stents was used in cross-bred piglet model. The procedure was performed under general anaesthesia with simulation of percutaneous approach. Open laparotomic view of duodenum was used for observation of AR valve.
Název v anglickém jazyce
Partially covered, self-expandable metal stents with anti-reflux valves for malignancies comprising hilar region
Popis výsledku anglicky
Objective: To prove feasibility of percutaneous parallel implantation of partially covered, self-expandable metal stents (PC-SEMS) with antireflux valves (AR valves) in in-vivo model. Method: The novel design of PC-SEMS suitable for patients with hilar malignancies was developed. For percutaneous implantation of stent with AR valve, specific type of introducer is needed to prevent closing and crumpling of valve during implantation procedure. Stent is half covered with e-PTFE membrane, also forming AR valve. Proximal half of the stent is bare to allow flow of bile from segmental branches of biliary tree arising in hilar region. In-vivo experiment was chosen to test safe deployment of two PC-SEMS with AR valve. Parallel and synchronous implantation of stents was used in cross-bred piglet model. The procedure was performed under general anaesthesia with simulation of percutaneous approach. Open laparotomic view of duodenum was used for observation of AR valve.
Klasifikace
Druh
O - Ostatní výsledky
CEP obor
FP - Ostatní lékařské obory
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2015