Amperometric biosensors for detection of phenol using chemically modified electrodes containing immobilized bacteria
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14310%2F02%3A00007179" target="_blank" >RIV/00216224:14310/02:00007179 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Amperometric biosensors for detection of phenol using chemically modified electrodes containing immobilized bacteria
Popis výsledku v původním jazyce
Eight strains of Pseudomonas were studied for development of phenol sensor. The immobilization of cells was performed by absorbing them on the working part of mediator-modified screen-printed electrodes (SPEs). Only three Pseudomonas strains were able totransfer electrons resulting from specific oxidation of phenol to the electrode by means of mediators; ferrocene, duroquinone and dimethyferrocene were successfully used with the strains 394 (p20), 74-III and 83-IV (working names), respectively. The lower limits for detection of phenol were 1 muM for the strain 74-III and 10 muM for the strain 83-IV and 394 (p20). Calibrations were obtained as the dependencies of logarithm of current changes (log DeltaI) on logarithm of concentration (log C), log DeltaI vs. log C. Among all substrates tested (phenol, catechol, hydroquinone, ethanol, methanol, propanol, isopropanol, isobutanol, isoamylalcohol, acetate, glucose, xylose, vanillin, 2,4,6-trichlorphenol, 2,3,6-trichlorphenol, 4-hydroxy-3-me
Název v anglickém jazyce
Amperometric biosensors for detection of phenol using chemically modified electrodes containing immobilized bacteria
Popis výsledku anglicky
Eight strains of Pseudomonas were studied for development of phenol sensor. The immobilization of cells was performed by absorbing them on the working part of mediator-modified screen-printed electrodes (SPEs). Only three Pseudomonas strains were able totransfer electrons resulting from specific oxidation of phenol to the electrode by means of mediators; ferrocene, duroquinone and dimethyferrocene were successfully used with the strains 394 (p20), 74-III and 83-IV (working names), respectively. The lower limits for detection of phenol were 1 muM for the strain 74-III and 10 muM for the strain 83-IV and 394 (p20). Calibrations were obtained as the dependencies of logarithm of current changes (log DeltaI) on logarithm of concentration (log C), log DeltaI vs. log C. Among all substrates tested (phenol, catechol, hydroquinone, ethanol, methanol, propanol, isopropanol, isobutanol, isoamylalcohol, acetate, glucose, xylose, vanillin, 2,4,6-trichlorphenol, 2,3,6-trichlorphenol, 4-hydroxy-3-me
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CE - Biochemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/OC%20518.30" target="_blank" >OC 518.30: Sledování interakcí biomakromolekul s biovrstvami v různých konformačních stavech pomocí piezoelektrických biosensorů</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2002
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Biosensors & Bioelectronics
ISSN
0956-5663
e-ISSN
—
Svazek periodika
17
Číslo periodika v rámci svazku
10
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
7
Strana od-do
867
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—