UV a IR laserová ablace organických práškových materiálů - příklad: ICP-OES s laserovou ablací spikovaných odlévaných tablet kojeneckého sušeného mléka
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14310%2F06%3A00018199" target="_blank" >RIV/00216224:14310/06:00018199 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
UV and IR laser ablation study of natural organic powdered materials?an example: ICP-OES with laser ablation of spiked infant formula cast pellets
Popis výsledku v původním jazyce
Laser ablation (LA) sample introduction into inductively coupled plasma optical emission spectrometry (ICP-OES) was used for powdered infant food formula being representative of samples with an organic matrix. Milk samples enriched with eight naturally occurring elements were presented for LA-ICP-OES experiments as cast epoxy resin disks. For Ca, Cu, Fe, Mg and Zn, a performance of IR (1064 nm) and UV (266 nm) ablation was compared in a fixed spot mode (UV) and in a scanning mode (both UV and IR). Signal vs. content dependences proved to be linear both with and w/o the internal standard Ca (IS). Using the IS the linear regression parameters for the scanning mode were improved for both UV and IR ablation. The UV fixed spot ablation provided the closestcorrelation and so, except for Cu (low content), repeatability for all elements of interest did not exceed 10% RSD. Lowest determinable quantities were 0.06% Ca, 0.008% Mg, 0.0005% Cu, 0.001% Fe and 0.0008% Zn. Condensation of Zn on eject
Název v anglickém jazyce
UV and IR laser ablation study of natural organic powdered materials?an example: ICP-OES with laser ablation of spiked infant formula cast pellets
Popis výsledku anglicky
Laser ablation (LA) sample introduction into inductively coupled plasma optical emission spectrometry (ICP-OES) was used for powdered infant food formula being representative of samples with an organic matrix. Milk samples enriched with eight naturally occurring elements were presented for LA-ICP-OES experiments as cast epoxy resin disks. For Ca, Cu, Fe, Mg and Zn, a performance of IR (1064 nm) and UV (266 nm) ablation was compared in a fixed spot mode (UV) and in a scanning mode (both UV and IR). Signal vs. content dependences proved to be linear both with and w/o the internal standard Ca (IS). Using the IS the linear regression parameters for the scanning mode were improved for both UV and IR ablation. The UV fixed spot ablation provided the closestcorrelation and so, except for Cu (low content), repeatability for all elements of interest did not exceed 10% RSD. Lowest determinable quantities were 0.06% Ca, 0.008% Mg, 0.0005% Cu, 0.001% Fe and 0.0008% Zn. Condensation of Zn on eject
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2006
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of analytical atomic spectrometry
ISSN
0267-9477
e-ISSN
—
Svazek periodika
21/2006
Číslo periodika v rámci svazku
9
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
4
Strana od-do
974-977