Srovnání double-pulse a single-pulse techniky spektroskopie laserem buzeného plazmatu v analýze práškových vzorků silikátových surovin pro průmysl keramiky a stavebních hmot
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14310%2F08%3A00026117" target="_blank" >RIV/00216224:14310/08:00026117 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparison of double-pulse and single-pulse laser-induced breakdown spectroscopy techniques in the analysis of powdered samples of silicate raw materials for brick-and-tile industry
Popis výsledku v původním jazyce
Single- and double-pulse laser-induced breakdown spectroscopy techniques applied to the analysis of pressed pellets of silicate raw materials were compared in terms of precision, sensitivity and limits of detection. Two Nd:YAG lasers (1064 and 532 mm) inan orthogonal configuration with a reheating arrangement have been employed. The main factors influencing system performance were optimized, i.e. laser pulse energies and interpulse separation. The behaviour of plasma temperature was studied over a period of time and calibration curves for Mg were constructed for both the single and double-pulse setup. When comparing the single- and double-pulse techniques, limits of detection of Si and Mg for the double-pulse technique were found to be 10 times lower.
Název v anglickém jazyce
Comparison of double-pulse and single-pulse laser-induced breakdown spectroscopy techniques in the analysis of powdered samples of silicate raw materials for brick-and-tile industry
Popis výsledku anglicky
Single- and double-pulse laser-induced breakdown spectroscopy techniques applied to the analysis of pressed pellets of silicate raw materials were compared in terms of precision, sensitivity and limits of detection. Two Nd:YAG lasers (1064 and 532 mm) inan orthogonal configuration with a reheating arrangement have been employed. The main factors influencing system performance were optimized, i.e. laser pulse energies and interpulse separation. The behaviour of plasma temperature was studied over a period of time and calibration curves for Mg were constructed for both the single and double-pulse setup. When comparing the single- and double-pulse techniques, limits of detection of Si and Mg for the double-pulse technique were found to be 10 times lower.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2008
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Spectrochimica Acta Part B: Atomic Spectroscopy
ISSN
0584-8547
e-ISSN
—
Svazek periodika
63
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
9
Strana od-do
—
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—