Universal Reference Electrode System for Non-Ideal Media
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14310%2F24%3A00138159" target="_blank" >RIV/00216224:14310/24:00138159 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61989592:15110/24:73627378
Výsledek na webu
<a href="http://www.chemicke-listy.cz/ojs3/index.php/chemicke-listy/article/view/4912/4873" target="_blank" >http://www.chemicke-listy.cz/ojs3/index.php/chemicke-listy/article/view/4912/4873</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.54779/chl20240602" target="_blank" >10.54779/chl20240602</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Universal Reference Electrode System for Non-Ideal Media
Popis výsledku v původním jazyce
Potentials in electrochemical experiments are referenced to a hydrogen electrode, which is set up in an ideal aqueous medium. However, most electrochemical experiments are performed in media which are far from ideal. This inevitably leads to the necessity of using a liquid junction between ideal and non-ideal solutions, or a pseudoreference electrode. The resulting undefined and irreproducible liquid junction potential can be eliminated by the Universal Reference Electrode system presented here. This consists of two electrodes: a voltammetric metallocene electrode and a potentiometric polyaniline electrode. The system is not a replacement for a Normal Hydrogen Electrode. It is only an elimination of undefined liquid junction potential, when working in a non-ideal medium.
Název v anglickém jazyce
Universal Reference Electrode System for Non-Ideal Media
Popis výsledku anglicky
Potentials in electrochemical experiments are referenced to a hydrogen electrode, which is set up in an ideal aqueous medium. However, most electrochemical experiments are performed in media which are far from ideal. This inevitably leads to the necessity of using a liquid junction between ideal and non-ideal solutions, or a pseudoreference electrode. The resulting undefined and irreproducible liquid junction potential can be eliminated by the Universal Reference Electrode system presented here. This consists of two electrodes: a voltammetric metallocene electrode and a potentiometric polyaniline electrode. The system is not a replacement for a Normal Hydrogen Electrode. It is only an elimination of undefined liquid junction potential, when working in a non-ideal medium.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10405 - Electrochemistry (dry cells, batteries, fuel cells, corrosion metals, electrolysis)
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2024
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
CHEMICKÉ LISTY
ISSN
0009-2770
e-ISSN
1213-7103
Svazek periodika
118
Číslo periodika v rámci svazku
11
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
6
Strana od-do
602-607
Kód UT WoS článku
001358436200005
EID výsledku v databázi Scopus
2-s2.0-85210259987