Time-of-Flight Based Calibration of an Ultrasonic Computed Tomography System
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14330%2F12%3A00080255" target="_blank" >RIV/00216224:14330/12:00080255 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216305:26220/12:PU95925
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Time-of-Flight Based Calibration of an Ultrasonic Computed Tomography System
Popis výsledku v původním jazyce
The paper presents a novel method for calibration of measuring geometry and of individual signal delays of transducers in ultrasonic computed tomography (USCT) systems via computational processing of multiple time-of-flight measurements of ultrasonic (US) impulses. The positions and time-delay parameters of thousands of ultrasonic transducers inside the USCT tank are calibrated by this approach with a high precision required for the tomographic reconstruction; such accuracy cannot be provided by any other known method. Although utilising similar basic principles as the global positioning system (GPS), the method is importantly generalised in treating all transducer parameters as the to-be calibrated (floating) unknowns, without any a-priori known positions and delays. The calibration is formulated as a non-linear least-squares problem, minimizing the differences between the calculated and measured time-of-arrivals of ultrasonic pulses.
Název v anglickém jazyce
Time-of-Flight Based Calibration of an Ultrasonic Computed Tomography System
Popis výsledku anglicky
The paper presents a novel method for calibration of measuring geometry and of individual signal delays of transducers in ultrasonic computed tomography (USCT) systems via computational processing of multiple time-of-flight measurements of ultrasonic (US) impulses. The positions and time-delay parameters of thousands of ultrasonic transducers inside the USCT tank are calibrated by this approach with a high precision required for the tomographic reconstruction; such accuracy cannot be provided by any other known method. Although utilising similar basic principles as the global positioning system (GPS), the method is importantly generalised in treating all transducer parameters as the to-be calibrated (floating) unknowns, without any a-priori known positions and delays. The calibration is formulated as a non-linear least-squares problem, minimizing the differences between the calculated and measured time-of-arrivals of ultrasonic pulses.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JC - Počítačový hardware a software
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/1M0572" target="_blank" >1M0572: Data, algoritmy, rozhodování</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radioengineering
ISSN
1210-2512
e-ISSN
—
Svazek periodika
21
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
12
Strana od-do
533-544
Kód UT WoS článku
000303135600023