Amperometric Immunosensor for Rapid Detection of Honeybee Pathogen Melissococcus Plutonius
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216224%3A14740%2F19%3A00108372" target="_blank" >RIV/00216224:14740/19:00108372 - isvavai.cz</a>
Výsledek na webu
<a href="https://onlinelibrary.wiley.com/doi/abs/10.1002/elan.201900252" target="_blank" >https://onlinelibrary.wiley.com/doi/abs/10.1002/elan.201900252</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1002/elan.201900252" target="_blank" >10.1002/elan.201900252</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Amperometric Immunosensor for Rapid Detection of Honeybee Pathogen Melissococcus Plutonius
Popis výsledku v původním jazyce
European foulbrood (EFB) is a honeybee larvae disease caused by a bacterium Melissococcus plutonius. An amperometric immunosensor based on a sandwich assay was developed for rapid point-of-care detection of this pathogen. An in-house made anti-Melissococcus antibody was immobilized to a gold surface of a screen-printed sensor via self-assembled monolayer of cysteamine activated with glutaraldehyde. The direct impedimetric detection of captured microbial cells was tested, however, a better performance was obtained after the formation of sandwich with the peroxidase-labeled antibody in the amperometric mode. The label-free assay was limited by higher non-specific binding. The limit of detection of the immunosensor was 6.6x10(4) CFU mL(-1) (colony-forming units) with wide linear range between 10(5) CFU mL(-1) and 10(9) CFU mL(-1). The whole analysis was completed within 2 h, which is shorter compared to common laboratory diagnostic tools, such as enzyme-linked immunosorbent assay or polymerase chain reaction. Furthermore, atomic force microscopy was used for confirmation of the bacteria presence on the electrode surface. The developed immunosensor was successfully employed in the analysis of real samples of honeybees and larvae. The achieved results demonstrate the potential of the amperometric immunosensor for practical in-field diagnosis of EFB, which can prevent infection spreading and connected losses of honeybee colonies.
Název v anglickém jazyce
Amperometric Immunosensor for Rapid Detection of Honeybee Pathogen Melissococcus Plutonius
Popis výsledku anglicky
European foulbrood (EFB) is a honeybee larvae disease caused by a bacterium Melissococcus plutonius. An amperometric immunosensor based on a sandwich assay was developed for rapid point-of-care detection of this pathogen. An in-house made anti-Melissococcus antibody was immobilized to a gold surface of a screen-printed sensor via self-assembled monolayer of cysteamine activated with glutaraldehyde. The direct impedimetric detection of captured microbial cells was tested, however, a better performance was obtained after the formation of sandwich with the peroxidase-labeled antibody in the amperometric mode. The label-free assay was limited by higher non-specific binding. The limit of detection of the immunosensor was 6.6x10(4) CFU mL(-1) (colony-forming units) with wide linear range between 10(5) CFU mL(-1) and 10(9) CFU mL(-1). The whole analysis was completed within 2 h, which is shorter compared to common laboratory diagnostic tools, such as enzyme-linked immunosorbent assay or polymerase chain reaction. Furthermore, atomic force microscopy was used for confirmation of the bacteria presence on the electrode surface. The developed immunosensor was successfully employed in the analysis of real samples of honeybees and larvae. The achieved results demonstrate the potential of the amperometric immunosensor for practical in-field diagnosis of EFB, which can prevent infection spreading and connected losses of honeybee colonies.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10406 - Analytical chemistry
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Electroanalysis
ISSN
1040-0397
e-ISSN
1521-4109
Svazek periodika
31
Číslo periodika v rámci svazku
10
Stát vydavatele periodika
DE - Spolková republika Německo
Počet stran výsledku
8
Strana od-do
1969-1976
Kód UT WoS článku
000474184300001
EID výsledku v databázi Scopus
2-s2.0-85068607764