Stanovení mezidetektorového objemu při SEC analýze polymerů
Popis výsledku
Byl stanoven interdetektorový objem při SEC chromatografii polymerů při použití MALS/koncentrační detektoru.
Klíčová slova
detectionLCmultiangle light scattering detectioninterdetector volumeradius of gyrationpeak broadeningpolymers
Identifikátory výsledku
Kód výsledku v IS VaVaI
Nalezeny alternativní kódy
RIV/61389013:_____/04:00105699
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Determination of the Interdetector Volume by s-detection in Size-Exclusion Chromatography of Polymers with on-line Multiangle Light-Scattering Detection
Popis výsledku v původním jazyce
The determination of the interdetector volume (IDV) in size-exclusion chromatography (SEC) of polymers with dual multiangle light-scattering/concentration detection based on the s-detection is demonstrated. By s-detection the determination is meant of local molecular mass from a known relation between molecular mass and radius of gyration, obtained from the angle dependence of the intensity of scattered light. IDV is found basing on the stipulation of equal slope of local calibrations found by the duallight-scattering/concentration detection, referred to as the w- and s-detection. As the s-detection uses just the light-scattering detector and, therefore, is independent of the value of IDV, this quantity can be found by searching for its value that gives the local calibration dependence obtained by the dual detection closest to that determined independently by the s-detection.
Název v anglickém jazyce
Determination of the Interdetector Volume by s-detection in Size-Exclusion Chromatography of Polymers with on-line Multiangle Light-Scattering Detection
Popis výsledku anglicky
The determination of the interdetector volume (IDV) in size-exclusion chromatography (SEC) of polymers with dual multiangle light-scattering/concentration detection based on the s-detection is demonstrated. By s-detection the determination is meant of local molecular mass from a known relation between molecular mass and radius of gyration, obtained from the angle dependence of the intensity of scattered light. IDV is found basing on the stipulation of equal slope of local calibrations found by the duallight-scattering/concentration detection, referred to as the w- and s-detection. As the s-detection uses just the light-scattering detector and, therefore, is independent of the value of IDV, this quantity can be found by searching for its value that gives the local calibration dependence obtained by the dual detection closest to that determined independently by the s-detection.
Klasifikace
Druh
Jx - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
KSK4050111: Cílená příprava chemických sloučenin s definovanými vlastnostmi
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2004
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Journal of Chromatography A
ISSN
0021-9673
e-ISSN
—
Svazek periodika
1045
Číslo periodika v rámci svazku
1-2
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
5
Strana od-do
37-41
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—
Základní informace
Druh výsledku
Jx - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP
CB - Analytická chemie, separace
Rok uplatnění
2004