Drilling efficiency and temperature elevation of three types of Kirschner-wire point
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26210%2F02%3APU36807" target="_blank" >RIV/00216305:26210/02:PU36807 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Drilling efficiency and temperature elevation of three types of Kirschner-wire point
Popis výsledku v původním jazyce
The paper deals with development of a new Kirschner wire (K-wire) design and its experimental testing in laboratory conditions and clinics for the use in orthopaedics, surgery, traumatology and clinics. The work contains general principles of Ilizarov external transosseous long bone fixation, bone drilling and pin implant insertion. Some mechanical engineering methods for the external skeletal implant insertion have been used, mainly based on insertion into polymethylmetacrylate, pig and goat bones (tibbias). Numerous tests of the K-wire performance tested in vitro and in vivo are described. The novel type of K-wire can be characterised by its superior position and statistically significant (Scheffé multiple range test, confidence level 95%) lowest effort to insertion, expressed in terms of axial force (35.23?1.43 N, torque moment 0.064?0.007 Nm, work needed for its insertion 32.6?2.5 J, and the lowest elevation of temperature (63?3°C) measured in the place of insertion compared to the
Název v anglickém jazyce
Drilling efficiency and temperature elevation of three types of Kirschner-wire point
Popis výsledku anglicky
The paper deals with development of a new Kirschner wire (K-wire) design and its experimental testing in laboratory conditions and clinics for the use in orthopaedics, surgery, traumatology and clinics. The work contains general principles of Ilizarov external transosseous long bone fixation, bone drilling and pin implant insertion. Some mechanical engineering methods for the external skeletal implant insertion have been used, mainly based on insertion into polymethylmetacrylate, pig and goat bones (tibbias). Numerous tests of the K-wire performance tested in vitro and in vivo are described. The novel type of K-wire can be characterised by its superior position and statistically significant (Scheffé multiple range test, confidence level 95%) lowest effort to insertion, expressed in terms of axial force (35.23?1.43 N, torque moment 0.064?0.007 Nm, work needed for its insertion 32.6?2.5 J, and the lowest elevation of temperature (63?3°C) measured in the place of insertion compared to the
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JQ - Strojní zařízení a nástroje
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA101%2F93%2F0569" target="_blank" >GA101/93/0569: Technologie vrtání a zevní transoseální skeletální fixace v ortopedii.</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2002
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
The Journal of Bone and Joint Surgery
ISSN
—
e-ISSN
—
Svazek periodika
84-B
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
4
Strana od-do
137-140
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—