Stress and reliability analyses of the hip joint endoprosthesis ceramic head with measured manufacturing inaccuracies
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26210%2F03%3APU37900" target="_blank" >RIV/00216305:26210/03:PU37900 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61388998:_____/03:51030128
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Stress and reliability analyses of the hip joint endoprosthesis ceramic head with measured manufacturing inaccuracies
Popis výsledku v původním jazyce
The paper presents the results of the computational modelling of the stress and reliability of the ceramic head of the total hip joint endoprosthesis under ISO 7206-5 loading. Endoprosthesis consists of the two main parts - ceramic head and the steel stem. The joint of these two parts is realised by the self-locking cone and the manufacturing inaccuracies of these cones significantly influence the tensile stresses in the head and head's failure probability. The manufacturing inaccuracies were measured oon 3D measuring device, stress in the analysed system is calculated using FEM system ANSYS and the head's reliability calculation is based in the Weibull weakest link theory.
Název v anglickém jazyce
Stress and reliability analyses of the hip joint endoprosthesis ceramic head with measured manufacturing inaccuracies
Popis výsledku anglicky
The paper presents the results of the computational modelling of the stress and reliability of the ceramic head of the total hip joint endoprosthesis under ISO 7206-5 loading. Endoprosthesis consists of the two main parts - ceramic head and the steel stem. The joint of these two parts is realised by the self-locking cone and the manufacturing inaccuracies of these cones significantly influence the tensile stresses in the head and head's failure probability. The manufacturing inaccuracies were measured oon 3D measuring device, stress in the analysed system is calculated using FEM system ANSYS and the head's reliability calculation is based in the Weibull weakest link theory.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
FS - Lékařská zařízení, přístroje a vybavení
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GP101%2F01%2FP039" target="_blank" >GP101/01/P039: Stochastický přístup k posouzení spolehlivosti keramických hlavic totální kyčelní endoprotézy</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2003
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Applied Mechanics 2003
ISBN
83-917224-3-0
ISSN
—
e-ISSN
—
Počet stran výsledku
6
Strana od-do
51-56
Název nakladatele
Politechnika Slaska, Katedra Mechaniky Stosowanej, Gliwice
Místo vydání
Jaworzynka, Poland
Místo konání akce
Jaworzynka
Datum konání akce
24. 3. 2003
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—