Stress analysis of the ceramic hip joint endoprosthesis with real shape deflections
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26210%2F04%3APU45022" target="_blank" >RIV/00216305:26210/04:PU45022 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61388998:_____/04:00103683
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Stress analysis of the ceramic hip joint endoprosthesis with real shape deflections
Popis výsledku v původním jazyce
The paper deals with the problems of ceramic head of hip joint endoprosthesis destructions, and with assessing the impact of shape deflections of conical surfaces on the probability of this failure. Concerned are shape deflections from the ideal conicalsurfaces of the stem and the head of the endoprosthesis, which - when the head is put on the stem and the endoprosthesis loaded - form a contact configuration. The shape deflections may be modelled at the macro-level - this concerns model shape inaccuraccies such as deflection from the nominal degree of taper, ovality, and their combination, or, possibly, at the micro-level - when the stochastic distribution of unevenness on the contact areas is respected. The problem of stress in ceramic heads was solved using the algorithm of the finite element method for spatial contact tasks, and the Weibull probability model was used for solving the problem of head cohesion failure probability. In the paper are presented and analysed the results of
Název v anglickém jazyce
Stress analysis of the ceramic hip joint endoprosthesis with real shape deflections
Popis výsledku anglicky
The paper deals with the problems of ceramic head of hip joint endoprosthesis destructions, and with assessing the impact of shape deflections of conical surfaces on the probability of this failure. Concerned are shape deflections from the ideal conicalsurfaces of the stem and the head of the endoprosthesis, which - when the head is put on the stem and the endoprosthesis loaded - form a contact configuration. The shape deflections may be modelled at the macro-level - this concerns model shape inaccuraccies such as deflection from the nominal degree of taper, ovality, and their combination, or, possibly, at the micro-level - when the stochastic distribution of unevenness on the contact areas is respected. The problem of stress in ceramic heads was solved using the algorithm of the finite element method for spatial contact tasks, and the Weibull probability model was used for solving the problem of head cohesion failure probability. In the paper are presented and analysed the results of
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
BO - Biofyzika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GP101%2F04%2FP037" target="_blank" >GP101/04/P037: Zvýšení úrovně výpočtového modelování pro určování pravděpodobnosti porušení keramické hlavice kyčelní endoprotézy</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2004
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Biomechanics of Man 2004
ISBN
80-7043-315-9
ISSN
—
e-ISSN
—
Počet stran výsledku
1
Strana od-do
107-107
Název nakladatele
University of West Bohemia
Místo vydání
Plzeň
Místo konání akce
Šumava Mountain - Hotel Horizont
Datum konání akce
16. 11. 2004
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—