In vitro measurement of pressure and flow characteristics of different nasal continuous positive airway pressure devices for neonates
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26210%2F17%3APU124002" target="_blank" >RIV/00216305:26210/17:PU124002 - isvavai.cz</a>
Výsledek na webu
<a href="http://www.jnpm.org/content/abstracts-25th-european-workshop-neonatology-westerburg-castle-germany-may-28th-%E2%80%93-31st-2017" target="_blank" >http://www.jnpm.org/content/abstracts-25th-european-workshop-neonatology-westerburg-castle-germany-may-28th-%E2%80%93-31st-2017</a>
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
In vitro measurement of pressure and flow characteristics of different nasal continuous positive airway pressure devices for neonates
Popis výsledku v původním jazyce
Our aim was to compare devices for nasal continuous positive airway pressure (NCPAP) application in neonates by measuring the pressure of the device and flow delivered to the patient. The following NCPAP delivery devices were tested: [Babyflow-Draeger (sizes XS, S, M, L, XL), Vygon single tube 2.5 mm, Vygon double tube 1.5 mm, RAM cannula (for newborns under 1.0 kg), Argyle pharyngeal cannula]. The generated flow was set to 8 liter/min (10 liter/min for L and XL), positive end-expiratory pressure was set to 5 mbar (6 mbar for L and XL).
Název v anglickém jazyce
In vitro measurement of pressure and flow characteristics of different nasal continuous positive airway pressure devices for neonates
Popis výsledku anglicky
Our aim was to compare devices for nasal continuous positive airway pressure (NCPAP) application in neonates by measuring the pressure of the device and flow delivered to the patient. The following NCPAP delivery devices were tested: [Babyflow-Draeger (sizes XS, S, M, L, XL), Vygon single tube 2.5 mm, Vygon double tube 1.5 mm, RAM cannula (for newborns under 1.0 kg), Argyle pharyngeal cannula]. The generated flow was set to 8 liter/min (10 liter/min for L and XL), positive end-expiratory pressure was set to 5 mbar (6 mbar for L and XL).
Klasifikace
Druh
O - Ostatní výsledky
CEP obor
—
OECD FORD obor
10305 - Fluids and plasma physics (including surface physics)
Návaznosti výsledku
Projekt
—
Návaznosti
S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů