In Situ Observation of Lubricant Film Formation in THR Considering Real Conformity: The Effect of Model Synovial Fluid Composition
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26210%2F18%3APU124594" target="_blank" >RIV/00216305:26210/18:PU124594 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/61989592:15110/18:73585626
Výsledek na webu
<a href="http://www.sciencedirect.com/science/article/pii/S0301679X17304164" target="_blank" >http://www.sciencedirect.com/science/article/pii/S0301679X17304164</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.triboint.2017.09.001" target="_blank" >10.1016/j.triboint.2017.09.001</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
In Situ Observation of Lubricant Film Formation in THR Considering Real Conformity: The Effect of Model Synovial Fluid Composition
Popis výsledku v původním jazyce
The present paper explores the effect of synovial fluid composition on lubricant film formation in hip replacements. The measurements were realized utilizing pendulum hip joint simulator, while the film thickness was evaluated using optical interferometry. Contact couples consisted of metal and ceramic femoral heads articulating with glass acetabular cups. As the test lubricants, various model fluids were employed. Initially, static tests, aimed on the effect of material and load on adsorption, were conducted. It was found that adsorbed film thickness increases independently of the head material. Consequently, swinging flexion-extension experiments were realized, revealing that the film formation is substantially affected by composition of model fluid. The thickest film was observed when higher concentration of hyaluronic acid and phospholipids was applied.
Název v anglickém jazyce
In Situ Observation of Lubricant Film Formation in THR Considering Real Conformity: The Effect of Model Synovial Fluid Composition
Popis výsledku anglicky
The present paper explores the effect of synovial fluid composition on lubricant film formation in hip replacements. The measurements were realized utilizing pendulum hip joint simulator, while the film thickness was evaluated using optical interferometry. Contact couples consisted of metal and ceramic femoral heads articulating with glass acetabular cups. As the test lubricants, various model fluids were employed. Initially, static tests, aimed on the effect of material and load on adsorption, were conducted. It was found that adsorbed film thickness increases independently of the head material. Consequently, swinging flexion-extension experiments were realized, revealing that the film formation is substantially affected by composition of model fluid. The thickest film was observed when higher concentration of hyaluronic acid and phospholipids was applied.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
20301 - Mechanical engineering
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2018
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Tribology International
ISSN
0301-679X
e-ISSN
1879-2464
Svazek periodika
117
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
11
Strana od-do
206-216
Kód UT WoS článku
000413880600022
EID výsledku v databázi Scopus
2-s2.0-85029360808