Zobecnený model v MRI zobrazení - transparentní cesta od RF signálu k obrazu
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F04%3APU40106" target="_blank" >RIV/00216305:26220/04:PU40106 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Generalised MRI signal model - a transparent way from RF signal to images
Popis výsledku v původním jazyce
The commonly used integral models of MRI signal are based on volume distribution of the static proton density as the signal source, sometimes modified by a multiplicative relaxation factor. By replacing this source function by the time-variable transversal magnetisation, a generalisation is obtained that enables to describe the MRI principles, different imaging techniques and image reconstruction methods in a transparent and consistent way. The generalisation enables to include all the relaxation phenoomena and differently weighted imaging into consideration. Also, the artefacts related to time-variance of the magnetisation distribution due to any other origin (e.g. flow and diffusion) can be included in the model as well. This way, even relatively complex phenomena like the time-variance of the magnetisation during data acquisition in fast MRI under complicated circumstances may in principle be included with a potential for deeper (perhaps numerical) analysis. Several examples of usin
Název v anglickém jazyce
Generalised MRI signal model - a transparent way from RF signal to images
Popis výsledku anglicky
The commonly used integral models of MRI signal are based on volume distribution of the static proton density as the signal source, sometimes modified by a multiplicative relaxation factor. By replacing this source function by the time-variable transversal magnetisation, a generalisation is obtained that enables to describe the MRI principles, different imaging techniques and image reconstruction methods in a transparent and consistent way. The generalisation enables to include all the relaxation phenoomena and differently weighted imaging into consideration. Also, the artefacts related to time-variance of the magnetisation distribution due to any other origin (e.g. flow and diffusion) can be included in the model as well. This way, even relatively complex phenomena like the time-variance of the magnetisation during data acquisition in fast MRI under complicated circumstances may in principle be included with a potential for deeper (perhaps numerical) analysis. Several examples of usin
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JD - Využití počítačů, robotika a její aplikace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA102%2F02%2F0890" target="_blank" >GA102/02/0890: Analýza medicínských ultrasonografických dat se zaměřením na 3D zobrazení v kardiologii</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2004
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proc. 17th International EURASIP conference BIOSIGNAL 2004
ISBN
80-214-2633-0
ISSN
—
e-ISSN
—
Počet stran výsledku
6
Strana od-do
1-6
Název nakladatele
VUTIUM Brno
Místo vydání
Brno
Místo konání akce
Brno
Datum konání akce
23. 6. 2004
Typ akce podle státní příslušnosti
EUR - Evropská akce