Gradient magnetic field measuring method using MR technique with spin echo series Měření gradientních magnetických polí metodou série spinových ech
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F04%3APU41370" target="_blank" >RIV/00216305:26220/04:PU41370 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68081731:_____/04:00100014
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Gradient magnetic field measuring method using MR technique with spin echo series
Popis výsledku v původním jazyce
An accurate method, suitable for measuring the time characteristics, is the method based on the MR, called the Instantaneous Frequency (IF) method [4]. The measurement can be performed on a commercial MR instrument with or without adjusted pre-emphasis compensation. This method can be used to find out the quality of the gradient magnetic fields with sufficient precision. A disadvantage of the IF method consist in the very limited time interval ? up to about 2.5 ms, when we are able to acquire the MR siggnal for the next processing. This was the reason for the basic IF method to be extended with spin echo and thus converted to the Instantaneous Frequency of Spin Echo (IFSE) method. The spin echo was created by using the p exciting pulse. The IFSE methodenables scanning the MR signal for an 18 ms time period. The third modification of the methods based on instantaneous frequency measurement is the Instantaneous Frequency of Spin Echoes Series (IFSES) method, which partially eliminates t
Název v anglickém jazyce
Gradient magnetic field measuring method using MR technique with spin echo series
Popis výsledku anglicky
An accurate method, suitable for measuring the time characteristics, is the method based on the MR, called the Instantaneous Frequency (IF) method [4]. The measurement can be performed on a commercial MR instrument with or without adjusted pre-emphasis compensation. This method can be used to find out the quality of the gradient magnetic fields with sufficient precision. A disadvantage of the IF method consist in the very limited time interval ? up to about 2.5 ms, when we are able to acquire the MR siggnal for the next processing. This was the reason for the basic IF method to be extended with spin echo and thus converted to the Instantaneous Frequency of Spin Echo (IFSE) method. The spin echo was created by using the p exciting pulse. The IFSE methodenables scanning the MR signal for an 18 ms time period. The third modification of the methods based on instantaneous frequency measurement is the Instantaneous Frequency of Spin Echoes Series (IFSES) method, which partially eliminates t
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/IAA2065201" target="_blank" >IAA2065201: Generace a měření gradientních magnetických polí pro in vivo MR lokalizovanou spektroskopii</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2004
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Progress in Electromagnetics Research
ISBN
88-8492-268-2
ISSN
—
e-ISSN
—
Počet stran výsledku
4
Strana od-do
1-4
Název nakladatele
University of Pisa
Místo vydání
Pisa
Místo konání akce
Pisa
Datum konání akce
28. 3. 2004
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—