Srovnání biosenzorových a chromatografických metod pro detekci pesticidů
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F05%3APU51991" target="_blank" >RIV/00216305:26220/05:PU51991 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/26234386:_____/05:#0000010
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparison of biosensoric and chromatographic methods for the detection of pesticides
Popis výsledku v původním jazyce
Screen-printed biosensors with immobilized acetylcholine esterase (AChE) were used for measuring fruit and vegetable samples that had first been analysed using gas and highperformance liquid chromatography. The output signal for the biosensors is the current, which is used to calculate relative inhibition (RI), a measured quantity. RI is proportional to the inhibiting (toxic) effect of organophosphates and carbamates. Measurements with AChE biosensors are not easily reproducible. This problem is soolvedby the choice of an arbitrary toxicity standard of 1.25 uM Syntostigmin. Measurements were evaluated by the ratio of the relative inhibition of the sample against the relative inhibition of Syntostigmin. Results obtained from the biosensor matchthose ofchromatography in 19 out of 38 total measurements made and for nine out of 19 positive samples. The confirmation rate was 50%. Future work must check the limit of 0.1 and the independent control of inhibiting pesticides contents after me
Název v anglickém jazyce
Comparison of biosensoric and chromatographic methods for the detection of pesticides
Popis výsledku anglicky
Screen-printed biosensors with immobilized acetylcholine esterase (AChE) were used for measuring fruit and vegetable samples that had first been analysed using gas and highperformance liquid chromatography. The output signal for the biosensors is the current, which is used to calculate relative inhibition (RI), a measured quantity. RI is proportional to the inhibiting (toxic) effect of organophosphates and carbamates. Measurements with AChE biosensors are not easily reproducible. This problem is soolvedby the choice of an arbitrary toxicity standard of 1.25 uM Syntostigmin. Measurements were evaluated by the ratio of the relative inhibition of the sample against the relative inhibition of Syntostigmin. Results obtained from the biosensor matchthose ofchromatography in 19 out of 38 total measurements made and for nine out of 19 positive samples. The confirmation rate was 50%. Future work must check the limit of 0.1 and the independent control of inhibiting pesticides contents after me
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2005
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
International Journal of Environmental Analytical Chemistry
ISSN
—
e-ISSN
—
Svazek periodika
85
Číslo periodika v rámci svazku
12-13
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
8
Strana od-do
885-892
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—