Gradient Coils Temperature Controller in a Gradient System for MR Tomography
Popis výsledku v původním jazyce
A simple gradient coils temperature controller for a gradient magnetic field system has been developed. In the application of modern fast magnetic resonance experiments it can maintain the temperature of gradient coils at a value of 21.2?C with a precision of 0.5?C and protect the gradient system against overloading by excessive short-circuit power. When the temperature and the maximum power are exceeded, it will disconnect all gradients and, simultaneously, signal this state to the operator. A gradientsystem complemented like this has been experimentally verified on a home-built 200 MHz Magnetic Resonance (MR) Tomography system with a 120 mm bore (with room temperature) in the Institute of Scientific Instruments, Academy of Sciences of the Czech Republic.
Název v anglickém jazyce
Gradient Coils Temperature Controller in a Gradient System for MR Tomography
Popis výsledku anglicky
A simple gradient coils temperature controller for a gradient magnetic field system has been developed. In the application of modern fast magnetic resonance experiments it can maintain the temperature of gradient coils at a value of 21.2?C with a precision of 0.5?C and protect the gradient system against overloading by excessive short-circuit power. When the temperature and the maximum power are exceeded, it will disconnect all gradients and, simultaneously, signal this state to the operator. A gradientsystem complemented like this has been experimentally verified on a home-built 200 MHz Magnetic Resonance (MR) Tomography system with a 120 mm bore (with room temperature) in the Institute of Scientific Instruments, Academy of Sciences of the Czech Republic.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GA102%2F07%2F0389" target="_blank" >GA102/07/0389: Výzkum nových NMR technik pro studium struktury porézních materiálů</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2007
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Reasearch in Telecomunication Technology 2007
ISBN
978-80-8070-735-4
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
1-5
Název nakladatele