Comparison of methods for absolute calibration of broadband AE sensors
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F12%3APU99803" target="_blank" >RIV/00216305:26220/12:PU99803 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparison of methods for absolute calibration of broadband AE sensors
Popis výsledku v původním jazyce
The paper contains a brief description of the current state in our calibration facility for the acoustic emission sensors. We have implemented the reciprocity method according to the NDIS 2109, ISO/TR 13115, using the step function according to ASTM 1106, ISO 12713 and the longitudinal wave calibration according to NPL. The greatest attention is given to problems relating to the correct implementation of AE sensor calibration methodology. It describes in detail influence of the aperture effect on the frequency response to the Rayleigh wave calibration of a wideband sensor. Briefly described is the basic principle and design of a wideband AE sensor with a conical active element and frequency characteristics of our reference sensors are specified as obtained from all types of absolute calibration for the surface waves and longitudinal waves.
Název v anglickém jazyce
Comparison of methods for absolute calibration of broadband AE sensors
Popis výsledku anglicky
The paper contains a brief description of the current state in our calibration facility for the acoustic emission sensors. We have implemented the reciprocity method according to the NDIS 2109, ISO/TR 13115, using the step function according to ASTM 1106, ISO 12713 and the longitudinal wave calibration according to NPL. The greatest attention is given to problems relating to the correct implementation of AE sensor calibration methodology. It describes in detail influence of the aperture effect on the frequency response to the Rayleigh wave calibration of a wideband sensor. Briefly described is the basic principle and design of a wideband AE sensor with a conical active element and frequency characteristics of our reference sensors are specified as obtained from all types of absolute calibration for the surface waves and longitudinal waves.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>S - Specificky vyzkum na vysokych skolach
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings 30th European Conference on Acoustic Emission Testing and 7th International Conference on Acoustic Emission
ISBN
978-84-615-9941-7
ISSN
—
e-ISSN
—
Počet stran výsledku
9
Strana od-do
"113-1"-"113-9"
Název nakladatele
University of Granada
Místo vydání
18071 Granada, Spain
Místo konání akce
Granada
Datum konání akce
12. 9. 2012
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—