The measurement of B0 and B1 fields inhomogeneity using the Spin Echo and Gradient The measurement of B0 and B1 fields inhomogeneity using the Spin Echo and Gradient Echo techniques
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F13%3APU104951" target="_blank" >RIV/00216305:26220/13:PU104951 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
The measurement of B0 and B1 fields inhomogeneity using the Spin Echo and Gradient The measurement of B0 and B1 fields inhomogeneity using the Spin Echo and Gradient Echo techniques
Popis výsledku v původním jazyce
This paper describes a method for the basic and RF field inhomogeneity estimation in NMR based on the measurement of T2 and T2*. In this paper a copper and plastic specimens is used in the Exponential approximation of the relaxation process. To distinguish the value of T2 and T2* relaxation times, we used the method both the gradient and spin echo acquisition techniques. A map of the spatial distribution of the B1 field is used to correct the inhomogeneity, and the quality of the B0 inhomogeneity strongly depends on successful approximation of T2. In this article the method of B1 estimation from the relaxation times via the magnetization transfer ratio is described.
Název v anglickém jazyce
The measurement of B0 and B1 fields inhomogeneity using the Spin Echo and Gradient The measurement of B0 and B1 fields inhomogeneity using the Spin Echo and Gradient Echo techniques
Popis výsledku anglicky
This paper describes a method for the basic and RF field inhomogeneity estimation in NMR based on the measurement of T2 and T2*. In this paper a copper and plastic specimens is used in the Exponential approximation of the relaxation process. To distinguish the value of T2 and T2* relaxation times, we used the method both the gradient and spin echo acquisition techniques. A map of the spatial distribution of the B1 field is used to correct the inhomogeneity, and the quality of the B0 inhomogeneity strongly depends on successful approximation of T2. In this article the method of B1 estimation from the relaxation times via the magnetization transfer ratio is described.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JA - Elektronika a optoelektronika, elektrotechnika
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
O - Projekt operacniho programu
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
International Interdisciplinary PhD Workshop 2013, Proceedings
ISBN
978-80-214-4759-2
ISSN
—
e-ISSN
—
Počet stran výsledku
5
Strana od-do
292-296
Název nakladatele
Neuveden
Místo vydání
Brno, Czech Republic
Místo konání akce
Brno
Datum konání akce
8. 9. 2013
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
—