Improving the Operating Parameters of the 300 kV AC Test System
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F19%3APU132157" target="_blank" >RIV/00216305:26220/19:PU132157 - isvavai.cz</a>
Výsledek na webu
<a href="https://ieeexplore.ieee.org/stamp/stamp.jsp?arnumber=8778099" target="_blank" >https://ieeexplore.ieee.org/stamp/stamp.jsp?arnumber=8778099</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1109/EPE.2019.8778099" target="_blank" >10.1109/EPE.2019.8778099</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Improving the Operating Parameters of the 300 kV AC Test System
Popis výsledku v původním jazyce
The High Voltage Test Laboratory in CVVOZEPowerLab is equipped with a Haefely Hipotronics HV Test Transformer HHT7300-300T5-760. This test system is used for withstand voltage, breakdown voltage tests and diagnostics of dielectrics - measurement of partial discharges and electronic bridge measurement up to 300 kV. Unfortunately, the system produces output voltage with high value of harmonic distortion, mainly in the condition of low output load. Higher harmonics in the output test voltage make it impossible to carry out accurate withstand tests and affects the measurement accuracy of the dissipation factor and the operation of the PD system. These problems were solved for last 5 years and currently are found several partial solutions. In this paper are described several basic concepts used to improve the operating parameters of the 300 kV AC source.
Název v anglickém jazyce
Improving the Operating Parameters of the 300 kV AC Test System
Popis výsledku anglicky
The High Voltage Test Laboratory in CVVOZEPowerLab is equipped with a Haefely Hipotronics HV Test Transformer HHT7300-300T5-760. This test system is used for withstand voltage, breakdown voltage tests and diagnostics of dielectrics - measurement of partial discharges and electronic bridge measurement up to 300 kV. Unfortunately, the system produces output voltage with high value of harmonic distortion, mainly in the condition of low output load. Higher harmonics in the output test voltage make it impossible to carry out accurate withstand tests and affects the measurement accuracy of the dissipation factor and the operation of the PD system. These problems were solved for last 5 years and currently are found several partial solutions. In this paper are described several basic concepts used to improve the operating parameters of the 300 kV AC source.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2019
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of the 2019 20th International Scientific Conference on Electric Power Engineering (EPE)
ISBN
978-1-7281-1333-3
ISSN
—
e-ISSN
—
Počet stran výsledku
6
Strana od-do
415-420
Název nakladatele
VSB - Technical University of Ostrava
Místo vydání
Ostrava
Místo konání akce
Kouty nad Desnou, Jeseniky mountains, Czech Repu
Datum konání akce
15. 5. 2019
Typ akce podle státní příslušnosti
WRD - Celosvětová akce
Kód UT WoS článku
000535890900080