Numerical modeling and experimental verification of a low fluid flow inductive flowmeter
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26220%2F21%3APU138445" target="_blank" >RIV/00216305:26220/21:PU138445 - isvavai.cz</a>
Výsledek na webu
<a href="http://apps.webofknowledge.com/full_record.do?product=WOS&search_mode=GeneralSearch&qid=26&SID=C1Ll2z1UrQnPSFmqh79&page=1&doc=1" target="_blank" >http://apps.webofknowledge.com/full_record.do?product=WOS&search_mode=GeneralSearch&qid=26&SID=C1Ll2z1UrQnPSFmqh79&page=1&doc=1</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.flowmeasinst.2020.101876" target="_blank" >10.1016/j.flowmeasinst.2020.101876</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Numerical modeling and experimental verification of a low fluid flow inductive flowmeter
Popis výsledku v původním jazyce
The article discusses modeling procedures to capture the physical and chemical processes present in operational measurement with a high-precision inductive flowmeter. In this context, a theoretical model and a numerical solution are proposed. Exploiting the combined finite element method (FEM) and the finite volume method (FVM), we prepared numerical models of multiple variants of the flowmeter and computed the output voltage on the device's electrodes. The model joins together the magnetic, electric, and current fields; a flow field; and a physical-chemical nonlinear ion model. The results were obtained by means of the FEM/FVM in ANSYS software and verified via experimental testing.
Název v anglickém jazyce
Numerical modeling and experimental verification of a low fluid flow inductive flowmeter
Popis výsledku anglicky
The article discusses modeling procedures to capture the physical and chemical processes present in operational measurement with a high-precision inductive flowmeter. In this context, a theoretical model and a numerical solution are proposed. Exploiting the combined finite element method (FEM) and the finite volume method (FVM), we prepared numerical models of multiple variants of the flowmeter and computed the output voltage on the device's electrodes. The model joins together the magnetic, electric, and current fields; a flow field; and a physical-chemical nonlinear ion model. The results were obtained by means of the FEM/FVM in ANSYS software and verified via experimental testing.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
20201 - Electrical and electronic engineering
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2021
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
FLOW MEASUREMENT AND INSTRUMENTATION
ISSN
0955-5986
e-ISSN
1873-6998
Svazek periodika
78
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
12
Strana od-do
1-12
Kód UT WoS článku
000632435200002
EID výsledku v databázi Scopus
—