Stanovení uranu pomocí ICP-OES při využití a neviužití prekoncentrace na makroporézních sorbentech
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26310%2F08%3APU76232" target="_blank" >RIV/00216305:26310/08:PU76232 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Determination of Uranium by ICP-AES in the Absence and Presence of Preconcentration on Macropourous Sorbents
Popis výsledku v původním jazyce
The echelle based ICP spectrometer at 27.12 MHz, with power output 1.35 kW, Meinhard concentric nebuliser and CID detector is used for the determination of U. Linear calibration plots for 0.1-50 ppm U(IV) are in solutions containing 1-3% HNO3 for 385.958and 409,014 nm. The slope of the plot decreases with the increasing concentration of HNO3 but increases in the presence of 4-(2-pyridylazo)resorcinol (PAR) by 10%. 3 x 10-3 M surfactants such as Septonex, Brij 35 and dodecylsulphate considerably increase the sensitivity of the determination of U. The detection limits are XD =1,26, XD =3,69 (Graham), XDm = 1,13 (Miller) or 0,30 ppm (IUPAC). No interferences at 1-3 ppm U were observed for 100:1 NO3-, Cl-, SO42-, SiO32- , NH4+, Na, K, Ca, Mg, Al 10:1 multicomponent containing Br-, Cl, Br-, SO42-, PO43-and Cd, Co, Cr, Cu, Mn, Pb, V, Zn in the absence or presence of internal standard Y (1ppm) except of Fe3+ for which the interfering signal 10:1 excess decreases only by 50% in the presence o
Název v anglickém jazyce
Determination of Uranium by ICP-AES in the Absence and Presence of Preconcentration on Macropourous Sorbents
Popis výsledku anglicky
The echelle based ICP spectrometer at 27.12 MHz, with power output 1.35 kW, Meinhard concentric nebuliser and CID detector is used for the determination of U. Linear calibration plots for 0.1-50 ppm U(IV) are in solutions containing 1-3% HNO3 for 385.958and 409,014 nm. The slope of the plot decreases with the increasing concentration of HNO3 but increases in the presence of 4-(2-pyridylazo)resorcinol (PAR) by 10%. 3 x 10-3 M surfactants such as Septonex, Brij 35 and dodecylsulphate considerably increase the sensitivity of the determination of U. The detection limits are XD =1,26, XD =3,69 (Graham), XDm = 1,13 (Miller) or 0,30 ppm (IUPAC). No interferences at 1-3 ppm U were observed for 100:1 NO3-, Cl-, SO42-, SiO32- , NH4+, Na, K, Ca, Mg, Al 10:1 multicomponent containing Br-, Cl, Br-, SO42-, PO43-and Cd, Co, Cr, Cu, Mn, Pb, V, Zn in the absence or presence of internal standard Y (1ppm) except of Fe3+ for which the interfering signal 10:1 excess decreases only by 50% in the presence o
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CB - Analytická chemie, separace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/OC%20183" target="_blank" >OC 183: Zavedení nových postupů pro monitorování xenobiotik ve vodách vypouštěných z vybraných čistíren odpadních vod ve městě Brně</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2008
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Chemické listy
ISSN
1213-7103
e-ISSN
—
Svazek periodika
102
Číslo periodika v rámci svazku
15
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
3
Strana od-do
—
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—