ELECTROCHEMICAL DETERMINATION OF LOW MOLECULAR WEIGHT ANTIOXIDANTS IN SERUM SAMPLES
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F00216305%3A26620%2F12%3APU126895" target="_blank" >RIV/00216305:26620/12:PU126895 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
ELECTROCHEMICAL DETERMINATION OF LOW MOLECULAR WEIGHT ANTIOXIDANTS IN SERUM SAMPLES
Popis výsledku v původním jazyce
The objective of the experiment was to establish and optimize an electrochemical method of low molecular weight antioxidants assay using square wave voltammetry performed on screen printed sensors with carbon working electrodes. The potential range of square wave voltammetry was 0 - 1.1 V. The optimal values for the assay, potential step 5 mV, amplitude 5 mV, frequency 1 Hz and volume of sample 20 µl, were found. The linearity and sensitivity of the method were verified by standard antioxidants - ascorbic acid, trolox, N-acetylcysteine and melatonin. The suitability for low molecular weight antioxidants` assay in human serum samples was confirmed by standard ferric reducing antioxidant power test (FRAP). Limits of detection were 0.3 mmol/l for ascorbic acid, 0.04 mmol/l for melatonin, 0.03 mmol/l for trolox, and 0.07 mmol/l for N-acetylcysteine. There was apparent a strong correlation between square wave voltammetry and the FRAP value as the correlation coefficient was 0.9778.
Název v anglickém jazyce
ELECTROCHEMICAL DETERMINATION OF LOW MOLECULAR WEIGHT ANTIOXIDANTS IN SERUM SAMPLES
Popis výsledku anglicky
The objective of the experiment was to establish and optimize an electrochemical method of low molecular weight antioxidants assay using square wave voltammetry performed on screen printed sensors with carbon working electrodes. The potential range of square wave voltammetry was 0 - 1.1 V. The optimal values for the assay, potential step 5 mV, amplitude 5 mV, frequency 1 Hz and volume of sample 20 µl, were found. The linearity and sensitivity of the method were verified by standard antioxidants - ascorbic acid, trolox, N-acetylcysteine and melatonin. The suitability for low molecular weight antioxidants` assay in human serum samples was confirmed by standard ferric reducing antioxidant power test (FRAP). Limits of detection were 0.3 mmol/l for ascorbic acid, 0.04 mmol/l for melatonin, 0.03 mmol/l for trolox, and 0.07 mmol/l for N-acetylcysteine. There was apparent a strong correlation between square wave voltammetry and the FRAP value as the correlation coefficient was 0.9778.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
—
OECD FORD obor
20601 - Medical engineering
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2012
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
MENDELNET 2012
ISBN
978-80-7375-836-3
ISSN
—
e-ISSN
—
Počet stran výsledku
8
Strana od-do
1225-1232
Název nakladatele
Neuveden
Místo vydání
Neuveden
Místo konání akce
Mendel Univ, Fac Agron, Brno
Datum konání akce
21. 11. 2012
Typ akce podle státní příslušnosti
CST - Celostátní akce
Kód UT WoS článku
000366461200143