Fricke-gel dosimetry in epithermal or thermal neutron beams of a research reactor
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F26722445%3A_____%2F15%3A%230001055" target="_blank" >RIV/26722445:_____/15:#0001055 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1016/j.radphyschem.2015.03.025" target="_blank" >http://dx.doi.org/10.1016/j.radphyschem.2015.03.025</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.radphyschem.2015.03.025" target="_blank" >10.1016/j.radphyschem.2015.03.025</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Fricke-gel dosimetry in epithermal or thermal neutron beams of a research reactor
Popis výsledku v původním jazyce
Fricke-xylenol-orange gel has shown noticeable potentiality for in-phantom dosimetry in epithermal or thermal neutron fields with very high fluence rate, as those characteristic of nuclear research reactors. Fricke gels in form of layers give the possibility of achieving spatial distribution of gamma dose, fast neutron dose and dose due to charged particles generated by thermal neutron reactions. The thermal neutron fluence has been deduced from the dose coming from the charge particles emitted by neutron reactions with the isotope B-10. Some measurements have been performed for improving the information on the relative sensitivity of Fricke gel dosimeters to the particles produced by B-10 reactions, because at present the precision of dose evaluations is limited by the scanty knowledge about the dependence of the dosimeter sensitivity on the radiation LET. For in-air measurements, the dosimeter material can produce an enhancement of thermal neutron fluence. Measurements and Monte Carlo calculations have been developed to investigate the importance of this effect.
Název v anglickém jazyce
Fricke-gel dosimetry in epithermal or thermal neutron beams of a research reactor
Popis výsledku anglicky
Fricke-xylenol-orange gel has shown noticeable potentiality for in-phantom dosimetry in epithermal or thermal neutron fields with very high fluence rate, as those characteristic of nuclear research reactors. Fricke gels in form of layers give the possibility of achieving spatial distribution of gamma dose, fast neutron dose and dose due to charged particles generated by thermal neutron reactions. The thermal neutron fluence has been deduced from the dose coming from the charge particles emitted by neutron reactions with the isotope B-10. Some measurements have been performed for improving the information on the relative sensitivity of Fricke gel dosimeters to the particles produced by B-10 reactions, because at present the precision of dose evaluations is limited by the scanty knowledge about the dependence of the dosimeter sensitivity on the radiation LET. For in-air measurements, the dosimeter material can produce an enhancement of thermal neutron fluence. Measurements and Monte Carlo calculations have been developed to investigate the importance of this effect.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JF - Jaderná energetika
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Radiation Physics and Chemistry
ISSN
0969-806X
e-ISSN
—
Svazek periodika
116
Číslo periodika v rámci svazku
November
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
7
Strana od-do
21-27
Kód UT WoS článku
000362606200005
EID výsledku v databázi Scopus
—