Gauge block calibration by means of the interferometry
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F46747885%3A24210%2F13%3A%230004473" target="_blank" >RIV/46747885:24210/13:#0004473 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00177016:_____/13:#0000817
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Gauge block calibration by means of the interferometry
Popis výsledku v původním jazyce
There are using two basic methods for gauge block calibration ? both are described by international standard EN ISO 3650. First of them is for calibration using laser, eventually multi-waves interferometry (interferential method). In this case gauge block (GB) is wrung to the reference plane from the free face of gauge block (gauge length). The second method is based on the comparison of length of calibrated GB with length of reference GB (comparative method). In this case there is used double-sided contact measurement by means of length incremental sensor. This paper deals with interferometry calibration method. Benefit of this method rest in its high accuracy. On the other hand it is time consuming method and there is also wear of gauge blocks. During measurement there was investigated GB surface from the microstructure point of view (microstructural analysis) by inerferometric method. It was proved than interferometric method has a great influence on the GB wear.
Název v anglickém jazyce
Gauge block calibration by means of the interferometry
Popis výsledku anglicky
There are using two basic methods for gauge block calibration ? both are described by international standard EN ISO 3650. First of them is for calibration using laser, eventually multi-waves interferometry (interferential method). In this case gauge block (GB) is wrung to the reference plane from the free face of gauge block (gauge length). The second method is based on the comparison of length of calibrated GB with length of reference GB (comparative method). In this case there is used double-sided contact measurement by means of length incremental sensor. This paper deals with interferometry calibration method. Benefit of this method rest in its high accuracy. On the other hand it is time consuming method and there is also wear of gauge blocks. During measurement there was investigated GB surface from the microstructure point of view (microstructural analysis) by inerferometric method. It was proved than interferometric method has a great influence on the GB wear.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/TA03010663" target="_blank" >TA03010663: Nové systémy pro kontrolu délky koncových měrek a vyhodnocení kvality jejich povrchů</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2013
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
International Conference ? TECHSYS 2013
ISBN
—
ISSN
1310-8271
e-ISSN
—
Počet stran výsledku
6
Strana od-do
231-236
Název nakladatele
Technical university - Sofia Branch Plovdiv
Místo vydání
Plovdiv ? Bulgaria
Místo konání akce
Plovdiv ? Bulgaria
Datum konání akce
1. 1. 2013
Typ akce podle státní příslušnosti
WRD - Celosvětová akce