A multiparametric heart failure score at baseline is associated with long-term outcome in patients with remotely monitored implantable cardioverter-defibrillators: A pooled analysis of 9 clinical trials
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F60076658%3A12110%2F25%3A43910435" target="_blank" >RIV/60076658:12110/25:43910435 - isvavai.cz</a>
Výsledek na webu
<a href="https://www.sciencedirect.com/science/article/pii/S1547527124034180?pes=vor&utm_source=scopus&getft_integrator=scopus" target="_blank" >https://www.sciencedirect.com/science/article/pii/S1547527124034180?pes=vor&utm_source=scopus&getft_integrator=scopus</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.hrthm.2024.10.005" target="_blank" >10.1016/j.hrthm.2024.10.005</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
A multiparametric heart failure score at baseline is associated with long-term outcome in patients with remotely monitored implantable cardioverter-defibrillators: A pooled analysis of 9 clinical trials
Popis výsledku v původním jazyce
BACKGROUND To predict worsening heart failure hospitalizations (WHFHs), the HeartInsight multiparametric algorithm calculates a heart failure (HF) Score based on temporal trends of physiologic parameters obtained through automatic daily remote monitoring of implantable cardioverter-defibrillators (ICDs). OBJECTIVE We studied the association of the baseline HF Score, determined at algorithm activation, with long-term patient outcomes. METHODS Data from 9 clinical trials were pooled, including 1841 ICD patients with a preimplantation ejection fraction <35%, New York Heart Association class II/III, and no long-standing atrial fibrillation. The primary end point was a composite of death or WHFH. RESULTS After a median follow-up of 631 days (interquartile range, 385-865 days), there were 243 WHFHs in 173 patients (9.4%) and 122 deaths (6.6%), 52 of which (42.6%) were cardiovascular. The primary end point occurred in 265 patients (14.4%). A multivariable time-to-first-event analysis showed that a high baseline HF Score (>23, as determined by a time-dependent receiver operating characteristics curve analysis) was significantly associated with the occurrence of the primary end point (adjusted hazard ratio [HR], 2.05; 95% confidence interval [CI], 1.54-2.71; P < .0001), all-cause death (HR, 2.37; CI, 1.56-3.58; P < .0001), cardiovascular death (HR, 2.19; CI, 1.14-4.22; P = .019), and WHFH (HR, 1.91; CI, 1.35-2.71; P = .0003). In a hierarchical event analysis of all-cause death as the outcome with highest priority and WHFHs as repeated event outcomes, the win ratio was 2.47 (CI, 1.89-3.24; P < .0001). CONCLUSION Based on a retrospective analysis of clinical trial data with adjudicated events, baseline HF Score derived from device-monitored variables was able to stratify patients at higher long-term risk of death or WHFH.
Název v anglickém jazyce
A multiparametric heart failure score at baseline is associated with long-term outcome in patients with remotely monitored implantable cardioverter-defibrillators: A pooled analysis of 9 clinical trials
Popis výsledku anglicky
BACKGROUND To predict worsening heart failure hospitalizations (WHFHs), the HeartInsight multiparametric algorithm calculates a heart failure (HF) Score based on temporal trends of physiologic parameters obtained through automatic daily remote monitoring of implantable cardioverter-defibrillators (ICDs). OBJECTIVE We studied the association of the baseline HF Score, determined at algorithm activation, with long-term patient outcomes. METHODS Data from 9 clinical trials were pooled, including 1841 ICD patients with a preimplantation ejection fraction <35%, New York Heart Association class II/III, and no long-standing atrial fibrillation. The primary end point was a composite of death or WHFH. RESULTS After a median follow-up of 631 days (interquartile range, 385-865 days), there were 243 WHFHs in 173 patients (9.4%) and 122 deaths (6.6%), 52 of which (42.6%) were cardiovascular. The primary end point occurred in 265 patients (14.4%). A multivariable time-to-first-event analysis showed that a high baseline HF Score (>23, as determined by a time-dependent receiver operating characteristics curve analysis) was significantly associated with the occurrence of the primary end point (adjusted hazard ratio [HR], 2.05; 95% confidence interval [CI], 1.54-2.71; P < .0001), all-cause death (HR, 2.37; CI, 1.56-3.58; P < .0001), cardiovascular death (HR, 2.19; CI, 1.14-4.22; P = .019), and WHFH (HR, 1.91; CI, 1.35-2.71; P = .0003). In a hierarchical event analysis of all-cause death as the outcome with highest priority and WHFHs as repeated event outcomes, the win ratio was 2.47 (CI, 1.89-3.24; P < .0001). CONCLUSION Based on a retrospective analysis of clinical trial data with adjudicated events, baseline HF Score derived from device-monitored variables was able to stratify patients at higher long-term risk of death or WHFH.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
30201 - Cardiac and Cardiovascular systems
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2025
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
HEART RHYTHM
ISSN
1547-5271
e-ISSN
1556-3871
Svazek periodika
22
Číslo periodika v rámci svazku
4
Stát vydavatele periodika
US - Spojené státy americké
Počet stran výsledku
9
Strana od-do
1055-1063
Kód UT WoS článku
001460684600001
EID výsledku v databázi Scopus
2-s2.0-85207715429