Real-time in-vivo mu-imaging with Medipix2
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F60076658%3A12310%2F09%3A00010512" target="_blank" >RIV/60076658:12310/09:00010512 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/68407700:21670/09:00165383
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Real-time in-vivo mu-imaging with Medipix2
Popis výsledku v původním jazyce
An X-ray micro-radiographic system based on the Medipix2 semiconductor pixel detector for dynamic high spatial resolution and for high contrast imaging has been developed. Our system is based on a micro-focus and nano-focus X-ray tube and the hybrid single-photon counting silicon pixel detector Medipix2 (matrix 256 x 256 sq. pixels of 55 mu m pitch). This compact table-top system stands promising as a new tool in the field of small animal imaging as well as in the in-vivo observation of dynamic processes inside living organisms. The main advantages of these Medipix2 pixel detectors include: high sensitivity to low-energy X-ray photons; position sensitive and noiseless single-photon detection with preselected photon energies: single-quantum counting ineach pixel performed by digital counter (therefore there is no dark current); digital integration (providing unlimited dynamic range and absolute linearity in device response to number of photons, high sensitivity and high contrast): real
Název v anglickém jazyce
Real-time in-vivo mu-imaging with Medipix2
Popis výsledku anglicky
An X-ray micro-radiographic system based on the Medipix2 semiconductor pixel detector for dynamic high spatial resolution and for high contrast imaging has been developed. Our system is based on a micro-focus and nano-focus X-ray tube and the hybrid single-photon counting silicon pixel detector Medipix2 (matrix 256 x 256 sq. pixels of 55 mu m pitch). This compact table-top system stands promising as a new tool in the field of small animal imaging as well as in the in-vivo observation of dynamic processes inside living organisms. The main advantages of these Medipix2 pixel detectors include: high sensitivity to low-energy X-ray photons; position sensitive and noiseless single-photon detection with preselected photon energies: single-quantum counting ineach pixel performed by digital counter (therefore there is no dark current); digital integration (providing unlimited dynamic range and absolute linearity in device response to number of photons, high sensitivity and high contrast): real
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
BG - Jaderná, atomová a molekulová fyzika, urychlovače
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Ostatní
Rok uplatnění
2009
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
ISSN
0168-9002
e-ISSN
—
Svazek periodika
607
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
3