Microradiography is an imaging technique using X-rays in the study of internal structures of objects. This rapid and convenient imaging tool is based on differential X-ray attenuation by various tissues and structures within the biological sample. The non-absorbed radiation is detected with a suitable detector and creates a radiographic image. In order to detect the differential properties of X-rays passing through structures sample with different compositions, an adequate high-quality imaging detectoris needed. We describe the recently developed radiographic apparatus, equipped with Timepix semiconductor pixel detector. The detector is used as an imager that counts individual photons of ionizing radiation, emitted by an X-ray tube FeinFocus with tungsten, copper or molybdenum anode. Thanks to the wide dynamic range, time over threshold mode-counter is used as Wilkinson type ADC allowing direct energy measurement in each pixel of Timepix detector and its high spatial resolution better
Název v anglickém jazyce
Microradiography of biological samples with Timepix
Popis výsledku anglicky
Microradiography is an imaging technique using X-rays in the study of internal structures of objects. This rapid and convenient imaging tool is based on differential X-ray attenuation by various tissues and structures within the biological sample. The non-absorbed radiation is detected with a suitable detector and creates a radiographic image. In order to detect the differential properties of X-rays passing through structures sample with different compositions, an adequate high-quality imaging detectoris needed. We describe the recently developed radiographic apparatus, equipped with Timepix semiconductor pixel detector. The detector is used as an imager that counts individual photons of ionizing radiation, emitted by an X-ray tube FeinFocus with tungsten, copper or molybdenum anode. Thanks to the wide dynamic range, time over threshold mode-counter is used as Wilkinson type ADC allowing direct energy measurement in each pixel of Timepix detector and its high spatial resolution better
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
EA - Morfologické obory a cytologie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
V - Vyzkumna aktivita podporovana z jinych verejnych zdroju
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
JOURNAL OF INSTRUMENTATION
ISSN
1748-0221
e-ISSN
—
Svazek periodika
6
Číslo periodika v rámci svazku
—
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
6
Strana od-do
1-6
Kód UT WoS článku
000298320400005
EID výsledku v databázi Scopus
—