Rapid determination of gross alpha/beta activity in urine with LSC
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F60162694%3AG45__%2F17%3A00533922" target="_blank" >RIV/60162694:G45__/17:00533922 - isvavai.cz</a>
Výsledek na webu
<a href="http://link.springer.com/article/10.1007/s10967-016-5077-x?wt_mc=Internal.Event.1.SEM.ArticleAuthorAssignedToIssue" target="_blank" >http://link.springer.com/article/10.1007/s10967-016-5077-x?wt_mc=Internal.Event.1.SEM.ArticleAuthorAssignedToIssue</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1007/s10967-016-5077-x" target="_blank" >10.1007/s10967-016-5077-x</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Rapid determination of gross alpha/beta activity in urine with LSC
Popis výsledku v původním jazyce
Rapid determination of gross alpha and beta emitters in urine by liquid scintillation counting is discussed. This method is based on direct addition of urine into scintillation cocktail. Am-241, Pu-239 and Sr-90 were selected as model radionuclides. The LSA Hidex 300 SL equipped with Triple-Double-Coincidence-Ratio technique was used for sample measurement. The work focused on optimizing the LSC cocktail to urine volume ratio with respect to the model radionuclides. The overall efficiencies for Am-241, Pu-239 and Sr-90 were greater than 92 %; therefore, this method would be suitable for rapid determination of gross alpha/beta activity.
Název v anglickém jazyce
Rapid determination of gross alpha/beta activity in urine with LSC
Popis výsledku anglicky
Rapid determination of gross alpha and beta emitters in urine by liquid scintillation counting is discussed. This method is based on direct addition of urine into scintillation cocktail. Am-241, Pu-239 and Sr-90 were selected as model radionuclides. The LSA Hidex 300 SL equipped with Triple-Double-Coincidence-Ratio technique was used for sample measurement. The work focused on optimizing the LSC cocktail to urine volume ratio with respect to the model radionuclides. The overall efficiencies for Am-241, Pu-239 and Sr-90 were greater than 92 %; therefore, this method would be suitable for rapid determination of gross alpha/beta activity.
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
50902 - Social sciences, interdisciplinary
Návaznosti výsledku
Projekt
—
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2017
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY
ISSN
0236-5731
e-ISSN
1588-2780
Svazek periodika
311
Číslo periodika v rámci svazku
1
Stát vydavatele periodika
CZ - Česká republika
Počet stran výsledku
5
Strana od-do
23-27
Kód UT WoS článku
000392075200003
EID výsledku v databázi Scopus
2-s2.0-84992332894