Comparative mechanical and corrosion studies on magnesium, zinc and iron alloys as biodegradable metals
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F60461373%3A22310%2F15%3A43900144" target="_blank" >RIV/60461373:22310/15:43900144 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Comparative mechanical and corrosion studies on magnesium, zinc and iron alloys as biodegradable metals
Popis výsledku v původním jazyce
In this paper, selected magnesium, zinc and iron biodegradable alloys were studied as prospective biomaterials for temporary medical implants like stents and fixation devices for fractured bones. Mechanical properties of the alloys were characterized with hardness and tensile tests. In-vitro corrosion behavior was studied using immersion tests in a simulated physiological solution (SPS, 9 g/L NaCl) to roughly estimate the in-vivo biodegradation rates of implants. It was found that the Mg and Zn alloys were limited by a tensile strength of 370 MPa, while the tensile strength of the Fe alloys achieved 530 MPa. The main advantage of the Mg alloys is that their Young's modulus of elasticity is similar to that of the human bone. However, the corrosion testsrevealed that the Mg-based alloys showed the highest corrosion rates in the SPS, ranging between 0.6 mm and 4.0 mm per year, which is above the tolerable degradation rates of implants. The corrosion rates of the Zn alloys were between 0.
Název v anglickém jazyce
Comparative mechanical and corrosion studies on magnesium, zinc and iron alloys as biodegradable metals
Popis výsledku anglicky
In this paper, selected magnesium, zinc and iron biodegradable alloys were studied as prospective biomaterials for temporary medical implants like stents and fixation devices for fractured bones. Mechanical properties of the alloys were characterized with hardness and tensile tests. In-vitro corrosion behavior was studied using immersion tests in a simulated physiological solution (SPS, 9 g/L NaCl) to roughly estimate the in-vivo biodegradation rates of implants. It was found that the Mg and Zn alloys were limited by a tensile strength of 370 MPa, while the tensile strength of the Fe alloys achieved 530 MPa. The main advantage of the Mg alloys is that their Young's modulus of elasticity is similar to that of the human bone. However, the corrosion testsrevealed that the Mg-based alloys showed the highest corrosion rates in the SPS, ranging between 0.6 mm and 4.0 mm per year, which is above the tolerable degradation rates of implants. The corrosion rates of the Zn alloys were between 0.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
JG - Hutnictví, kovové materiály
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GBP108%2F12%2FG043" target="_blank" >GBP108/12/G043: Mikro- a nanokrystalické materiály s vysokým podílem rozhraní pro moderní strukturní aplikace, biodegradabilní implantáty a uchovávání vodíku</a><br>
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)
Ostatní
Rok uplatnění
2015
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Materiali In Technologie
ISSN
1580-2949
e-ISSN
—
Svazek periodika
49
Číslo periodika v rámci svazku
6
Stát vydavatele periodika
SI - Slovinská republika
Počet stran výsledku
6
Strana od-do
877-882
Kód UT WoS článku
000366552500005
EID výsledku v databázi Scopus
—