Modelování reakčně transportních procesů v mikrokapilárním biosenzoru pro detekci lidských IgG
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F60461373%3A22340%2F06%3A00016652" target="_blank" >RIV/60461373:22340/06:00016652 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Modeling reaction-transport processes in a microcapillary biosensor for detection of human IgG
Popis výsledku v původním jazyce
Mathematical modeling of electrokinetic transport processes and antigen-antibody interactions in a microfluidic biosensor for heterogeneous immunoassay is described. Two mathematical models of externally controlled electroosmotic transport are constructed - the slip model based on Helmholtz-Smoluchowski approximation and the non-slip model based on the use of Poisson equation. It was found that the use of the slip model is restricted by concentration of the carrier electrolyte and dimensions of the system. The antigen-antibody interaction cannot be described by the slip model. Dynamics of the heterogeneous immunoassay in the proposed biosensor was also studied. Human-IgG-Protein A was used as model system. Parametric studies revealed strong sensitivityof the biosensor to the antibody charge number and to the character of surface heterogeneity in the position of immobilized antigen. This sensitivity can be one of the reasons why the practical applications of heterogeneous protein chips
Název v anglickém jazyce
Modeling reaction-transport processes in a microcapillary biosensor for detection of human IgG
Popis výsledku anglicky
Mathematical modeling of electrokinetic transport processes and antigen-antibody interactions in a microfluidic biosensor for heterogeneous immunoassay is described. Two mathematical models of externally controlled electroosmotic transport are constructed - the slip model based on Helmholtz-Smoluchowski approximation and the non-slip model based on the use of Poisson equation. It was found that the use of the slip model is restricted by concentration of the carrier electrolyte and dimensions of the system. The antigen-antibody interaction cannot be described by the slip model. Dynamics of the heterogeneous immunoassay in the proposed biosensor was also studied. Human-IgG-Protein A was used as model system. Parametric studies revealed strong sensitivityof the biosensor to the antibody charge number and to the character of surface heterogeneity in the position of immobilized antigen. This sensitivity can be one of the reasons why the practical applications of heterogeneous protein chips
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CI - Průmyslová chemie a chemické inženýrství
OECD FORD obor
—
Návaznosti výsledku
Projekt
<a href="/cs/project/GP104%2F03%2FD006" target="_blank" >GP104/03/D006: Studium mikroreaktorů pro bioinženýrské aplikace</a><br>
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2006
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Microelectronic Engineering
ISSN
0167-9317
e-ISSN
—
Svazek periodika
83
Číslo periodika v rámci svazku
4-9
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
4
Strana od-do
1660-1663
Kód UT WoS článku
—
EID výsledku v databázi Scopus
—