Measurement of atmospheric pollutants concentration
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F60461373%3A22340%2F10%3A00023437" target="_blank" >RIV/60461373:22340/10:00023437 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Measurement of atmospheric pollutants concentration
Popis výsledku v původním jazyce
A new laboratory station for measurement of concentration of atmospheric pollutants was constructed at the ICT Prague. The basis of the station is a system of automatic gas analyzers Advance Optima, which consists of infrared analyzing modules URAS 14 (measurement of the CO2, CO, CH4), thermomagnetic module MAGNOS 16 (measurement of the O2) and central unit. The analyzing system takes a sample of flue gases from the combustion chamber by a cooling unit Advance SCC with a sample pump. The combustion chamber consists of a burner with a spark gap and a flame ionization sensor. Flame in the furnace is initiated and maintained using a standard furnace automatic. In the combustion chamber the natural gas is burned with air. The exact ratio of mixture naturalgas/air is set by two controllers for the flow measurement of gaseous media, which are controlled by an industrial PC. Software for the station was programmed in LabVIEW.
Název v anglickém jazyce
Measurement of atmospheric pollutants concentration
Popis výsledku anglicky
A new laboratory station for measurement of concentration of atmospheric pollutants was constructed at the ICT Prague. The basis of the station is a system of automatic gas analyzers Advance Optima, which consists of infrared analyzing modules URAS 14 (measurement of the CO2, CO, CH4), thermomagnetic module MAGNOS 16 (measurement of the O2) and central unit. The analyzing system takes a sample of flue gases from the combustion chamber by a cooling unit Advance SCC with a sample pump. The combustion chamber consists of a burner with a spark gap and a flame ionization sensor. Flame in the furnace is initiated and maintained using a standard furnace automatic. In the combustion chamber the natural gas is burned with air. The exact ratio of mixture naturalgas/air is set by two controllers for the flow measurement of gaseous media, which are controlled by an industrial PC. Software for the station was programmed in LabVIEW.
Klasifikace
Druh
D - Stať ve sborníku
CEP obor
JB - Senzory, čidla, měření a regulace
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2010
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název statě ve sborníku
Proceedings of 9th International Conference Process Control 2010
ISBN
978-80-7399-951-3
ISSN
—
e-ISSN
—
Počet stran výsledku
1
Strana od-do
—
Název nakladatele
Tribun EU, s.r.o.
Místo vydání
Brno
Místo konání akce
Kouty nad Desnou
Datum konání akce
1. 1. 2010
Typ akce podle státní příslušnosti
EUR - Evropská akce
Kód UT WoS článku
—