Quantification of Volatile Compounds in the Headspace of Aqueous Liguids Using Selected Ion Flow Tube Mass Spectrometry.
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F02%3A54020242" target="_blank" >RIV/61388955:_____/02:54020242 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Quantification of Volatile Compounds in the Headspace of Aqueous Liguids Using Selected Ion Flow Tube Mass Spectrometry.
Popis výsledku v původním jazyce
We describe a method by which the concentrations of volatile compounds in the headspace of their dilute aqueous solutions in sealed containers can be determined using on-line selected ion flow tube mass spectrometry, SIFT-MS. Thus, the changing number density of the molecules of the volatile compound in the carrier gas of the SIFT-MS instrument is described in terms of its changing flow rate as the pressure in the sealed container decreases during the sampling procedure. It is shown that the best analytical procedure is to determine the mean concentration of the trace gas in the liquid headspace over a given sampling time and relate this to the required concentration, which is the initial equilibrium concentration established before the pressure in thesealed container reduces significantly. The HenryŽs Law constants for these compounds have been determined and found to agree with the published values.
Název v anglickém jazyce
Quantification of Volatile Compounds in the Headspace of Aqueous Liguids Using Selected Ion Flow Tube Mass Spectrometry.
Popis výsledku anglicky
We describe a method by which the concentrations of volatile compounds in the headspace of their dilute aqueous solutions in sealed containers can be determined using on-line selected ion flow tube mass spectrometry, SIFT-MS. Thus, the changing number density of the molecules of the volatile compound in the carrier gas of the SIFT-MS instrument is described in terms of its changing flow rate as the pressure in the sealed container decreases during the sampling procedure. It is shown that the best analytical procedure is to determine the mean concentration of the trace gas in the liquid headspace over a given sampling time and relate this to the required concentration, which is the initial equilibrium concentration established before the pressure in thesealed container reduces significantly. The HenryŽs Law constants for these compounds have been determined and found to agree with the published values.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CF - Fyzikální chemie a teoretická chemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
P - Projekt vyzkumu a vyvoje financovany z verejnych zdroju (s odkazem do CEP)<br>Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2002
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Rapid Communications in Mass Spectrometry
ISSN
0951-4198
e-ISSN
—
Svazek periodika
16
Číslo periodika v rámci svazku
N/A
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
6
Strana od-do
2148-2153