The quantification of carbon dioxide in humid air and exhaled breath by selected ion flow tube mass spectrometry
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F09%3A00329750" target="_blank" >RIV/61388955:_____/09:00329750 - isvavai.cz</a>
Výsledek na webu
—
DOI - Digital Object Identifier
—
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
The quantification of carbon dioxide in humid air and exhaled breath by selected ion flow tube mass spectrometry
Popis výsledku v původním jazyce
The reactions of carbon dioxide, CO2, with the precursor ions used for selected ion flow tube mass spectrometry, SIFT-MS, analyses, viz. H3O+, NO+ and O2+, are so slow that the presence of CO2 in exhaled breath has, until recently, not had to be accounted for in SIFT-MS analyses of breath. This has, however, to be accounted for in the analysis of acetaldehyde in breath, because an overlap occurs of the monohydrate of protonated acetaldehyde and the weakly bound adduct ion, H3O+CO2, formed by the slow association reaction of the precursor ion H3O+ with CO2 molecules. The understanding of the kinetics of formation and the loss rates of the relevant ions gained from experimentation using the new generation of more sensitive SIFT-MS instruments now allowsaccurate quantification of CO2 in breath using the level of the H3O+CO2 adduct ion. However, this is complicated by the rapid reaction of H3O+CO2 with water vapour molecules, H2O, that are in abundance in exhaled breath.
Název v anglickém jazyce
The quantification of carbon dioxide in humid air and exhaled breath by selected ion flow tube mass spectrometry
Popis výsledku anglicky
The reactions of carbon dioxide, CO2, with the precursor ions used for selected ion flow tube mass spectrometry, SIFT-MS, analyses, viz. H3O+, NO+ and O2+, are so slow that the presence of CO2 in exhaled breath has, until recently, not had to be accounted for in SIFT-MS analyses of breath. This has, however, to be accounted for in the analysis of acetaldehyde in breath, because an overlap occurs of the monohydrate of protonated acetaldehyde and the weakly bound adduct ion, H3O+CO2, formed by the slow association reaction of the precursor ion H3O+ with CO2 molecules. The understanding of the kinetics of formation and the loss rates of the relevant ions gained from experimentation using the new generation of more sensitive SIFT-MS instruments now allowsaccurate quantification of CO2 in breath using the level of the H3O+CO2 adduct ion. However, this is complicated by the rapid reaction of H3O+CO2 with water vapour molecules, H2O, that are in abundance in exhaled breath.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CF - Fyzikální chemie a teoretická chemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
Výsledek vznikl pri realizaci vícero projektů. Více informací v záložce Projekty.
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2009
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Rapid Communications in Mass Spectrometry
ISSN
0951-4198
e-ISSN
—
Svazek periodika
23
Číslo periodika v rámci svazku
10
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
7
Strana od-do
—
Kód UT WoS článku
000265985000003
EID výsledku v databázi Scopus
—