On-line, real time monitoring of exhaled trace gases by SIFT-MS in the perioperative setting: a feasibility study
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F11%3A00363284" target="_blank" >RIV/61388955:_____/11:00363284 - isvavai.cz</a>
Výsledek na webu
<a href="http://dx.doi.org/10.1039/C1AN15356K" target="_blank" >http://dx.doi.org/10.1039/C1AN15356K</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1039/C1AN15356K" target="_blank" >10.1039/C1AN15356K</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
On-line, real time monitoring of exhaled trace gases by SIFT-MS in the perioperative setting: a feasibility study
Popis výsledku v původním jazyce
A study is described of the first on line, real time analyses of the exhaled breath of five anaesthetized patients during the complete perioperative periods of laparoscopic surgery. These breath analyses were achieved using a selected ion flow tube, SIFT-MS, instrument, located in the operating theatre at an acceptable distance from the operating table, and coupled to the endotracheal tube in the ventilation circuit via a 5 metre long capillary tube. Thus, inhalation/exhalation breathing cycles, set tobe at a frequency of 10 per minute, were sampled continuously for water vapour, the metabolites acetone and isoprene and the propofol used to induce anaesthesia for each operating period that ranged from 20 min (shortest) to 80 min (longest). Whilst there was some loss of water vapour along the long sampling line, the concentrations of the other trace compounds were not diminished.
Název v anglickém jazyce
On-line, real time monitoring of exhaled trace gases by SIFT-MS in the perioperative setting: a feasibility study
Popis výsledku anglicky
A study is described of the first on line, real time analyses of the exhaled breath of five anaesthetized patients during the complete perioperative periods of laparoscopic surgery. These breath analyses were achieved using a selected ion flow tube, SIFT-MS, instrument, located in the operating theatre at an acceptable distance from the operating table, and coupled to the endotracheal tube in the ventilation circuit via a 5 metre long capillary tube. Thus, inhalation/exhalation breathing cycles, set tobe at a frequency of 10 per minute, were sampled continuously for water vapour, the metabolites acetone and isoprene and the propofol used to induce anaesthesia for each operating period that ranged from 20 min (shortest) to 80 min (longest). Whilst there was some loss of water vapour along the long sampling line, the concentrations of the other trace compounds were not diminished.
Klasifikace
Druh
J<sub>x</sub> - Nezařazeno - Článek v odborném periodiku (Jimp, Jsc a Jost)
CEP obor
CF - Fyzikální chemie a teoretická chemie
OECD FORD obor
—
Návaznosti výsledku
Projekt
—
Návaznosti
Z - Vyzkumny zamer (s odkazem do CEZ)
Ostatní
Rok uplatnění
2011
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Analyst
ISSN
0003-2654
e-ISSN
—
Svazek periodika
136
Číslo periodika v rámci svazku
16
Stát vydavatele periodika
GB - Spojené království Velké Británie a Severního Irska
Počet stran výsledku
5
Strana od-do
3233-3237
Kód UT WoS článku
000293085000002
EID výsledku v databázi Scopus
—