Differential pulse voltammetric determination of homovanillic acid as a tumor biomarker in human urine after hollow fiber-based liquid-phase microextraction
Identifikátory výsledku
Kód výsledku v IS VaVaI
<a href="https://www.isvavai.cz/riv?ss=detail&h=RIV%2F61388955%3A_____%2F21%3A00532424" target="_blank" >RIV/61388955:_____/21:00532424 - isvavai.cz</a>
Nalezeny alternativní kódy
RIV/00216208:11310/21:10417793
Výsledek na webu
<a href="http://hdl.handle.net/11104/0310920" target="_blank" >http://hdl.handle.net/11104/0310920</a>
DOI - Digital Object Identifier
<a href="http://dx.doi.org/10.1016/j.talanta.2020.121594" target="_blank" >10.1016/j.talanta.2020.121594</a>
Alternativní jazyky
Jazyk výsledku
angličtina
Název v původním jazyce
Differential pulse voltammetric determination of homovanillic acid as a tumor biomarker in human urine after hollow fiber-based liquid-phase microextraction
Popis výsledku v původním jazyce
Novel method for the determination of a tumor marker homovanillic acid (HVA) in human urine was developed. Combination of hollow fiber – based liquid-phase microextraction (HF-LPME) and differential pulse voltammetry (DPV) at a cathodically pre-treated boron doped diamond electrode (BDDE) was applied for these purposes. Optimum conditions were: butyl benzoate as supported liquid membrane (SLM) formed on polypropylene HF, 0.1 mol L−1 HCl as donor phase, 0.1 mol L−1 sodium phosphate buffer of pH 6 as acceptor phase, and 30 min extraction time. HF-LPME-DPV concentration dependence was linear in the range from 1.2 to 100 μmol L−1. Limits of quantification (LOQ) and detection (LOD) were 1.2 and 0.4 μmol L−1, respectively. The applicability of the developed method was verified by analysis of human urine. Standard addition method was used, found HVA concentration was 13.5 ± 1.3 μmol L−1, RSD = 9.3% (n=5).
Název v anglickém jazyce
Differential pulse voltammetric determination of homovanillic acid as a tumor biomarker in human urine after hollow fiber-based liquid-phase microextraction
Popis výsledku anglicky
Novel method for the determination of a tumor marker homovanillic acid (HVA) in human urine was developed. Combination of hollow fiber – based liquid-phase microextraction (HF-LPME) and differential pulse voltammetry (DPV) at a cathodically pre-treated boron doped diamond electrode (BDDE) was applied for these purposes. Optimum conditions were: butyl benzoate as supported liquid membrane (SLM) formed on polypropylene HF, 0.1 mol L−1 HCl as donor phase, 0.1 mol L−1 sodium phosphate buffer of pH 6 as acceptor phase, and 30 min extraction time. HF-LPME-DPV concentration dependence was linear in the range from 1.2 to 100 μmol L−1. Limits of quantification (LOQ) and detection (LOD) were 1.2 and 0.4 μmol L−1, respectively. The applicability of the developed method was verified by analysis of human urine. Standard addition method was used, found HVA concentration was 13.5 ± 1.3 μmol L−1, RSD = 9.3% (n=5).
Klasifikace
Druh
J<sub>imp</sub> - Článek v periodiku v databázi Web of Science
CEP obor
—
OECD FORD obor
10405 - Electrochemistry (dry cells, batteries, fuel cells, corrosion metals, electrolysis)
Návaznosti výsledku
Projekt
<a href="/cs/project/GA20-01589S" target="_blank" >GA20-01589S: Nové strategie pro zlepšení senzorových vlastností nových elektrodových materiálů prostřednictvím jejich předúpravy či modifikace povrchu</a><br>
Návaznosti
I - Institucionalni podpora na dlouhodoby koncepcni rozvoj vyzkumne organizace
Ostatní
Rok uplatnění
2021
Kód důvěrnosti údajů
S - Úplné a pravdivé údaje o projektu nepodléhají ochraně podle zvláštních právních předpisů
Údaje specifické pro druh výsledku
Název periodika
Talanta
ISSN
0039-9140
e-ISSN
1873-3573
Svazek periodika
221
Číslo periodika v rámci svazku
JAN 2021
Stát vydavatele periodika
NL - Nizozemsko
Počet stran výsledku
7
Strana od-do
121594
Kód UT WoS článku
000579910600084
EID výsledku v databázi Scopus
2-s2.0-85091253335